A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Automated external defibrillators on board merchant vessels? Preliminary report article for discussion. | LitMetric

Automated external defibrillators on board merchant vessels? Preliminary report article for discussion.

Int Marit Health

See-Berufsgenossenschaft, Reimerswiete 2, D-20457 Hamburg, Germany.

Published: July 2006

Objectives: Acute heart diseases are the most frequent causes for fatalities on merchant vessels. Presently there is no sufficient therapy available to treat ventricular fibrillation. The aim of this study was to test whether common automated external defibrillators [AED] may be appropriate for the use aboard merchant vessels.

Methods: In 2005, nine seafarers were introduced to four common models of AED (HeartStartFR2+, Lifepak500, AEDplus, FREDeasy) using standard video or DVD presentations. AED handling by the subjects was tested in standardized simulated emergency scenarios. After training, they subjectively rated each AED on 24 factors involved in the introduction and handling of the device. An actual ECG was then obtained with each AED at a site located beside the ship's main engine to test under maximum vibration. The ECG data were extracted and sent as an e-mail attachment via satellite to the German Telemedical Maritime Assistance Service [TMAS] in Cuxhaven.

Results And Conclusions: All subjects handled the AED correctly. The AED received a total amount of points in the range between 2125 to 2241 (of 2400 possible). The subjects preferred AED with coloured as well as light marked buttons which gave a feedback (e.g. audible tones) when they were pressed. All AED were able to register an ECG in the vibrating ambient. Due to interface problems it was only possible to extract three ECG files, and only two files (data < 300 kB) could be sent as e-mail attachment via satellite to the German TMAS. In noisy areas the AED must guide the user, e.g. by screen massages and/or pictograms. Displays should provide additional data to help assess resuscitation effectiveness. A special procedure is necessary to ensure that ships and TMAS own the same software to read the transmitted ECG files, which are not allowed to exceed a size of 300 kB.

Download full-text PDF

Source

Publication Analysis

Top Keywords

aed
9
automated external
8
external defibrillators
8
e-mail attachment
8
attachment satellite
8
satellite german
8
ecg files
8
ecg
5
defibrillators board
4
board merchant
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!