This study quantified the torsional resistance provided by locking plates and angled blade plates used to stabilize proximal humeral fractures. Three-part proximal humeral fractures were created in 6 pairs of cadaveric humeri. One specimen of each pair was reconstructed with a proximal humeral locking plate, whereas the other specimen was reconstructed with an angled blade plate. An external rotation torque, varying from 0 to 5 N-m, was applied to the humeral head until the head rotated 30 degrees or 10,000 loading cycles were applied. The mean initial torsional stiffness was significantly larger for the locking plates (0.99 N-m/degree) than for the blade plates (0.59 N-m/degree). For each pair, the maximum rotation was larger for the blade plate than for the locking plate. For this in vitro model of a reconstructed 3-part proximal humeral fracture, the locking plate provided better torsional fatigue resistance and stiffness than the blade plate.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jse.2005.08.006DOI Listing

Publication Analysis

Top Keywords

proximal humeral
20
locking plates
12
humeral fractures
12
locking plate
12
blade plate
12
torsional resistance
8
three-part proximal
8
angled blade
8
blade plates
8
locking
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!