Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as well as the automated count, included metamyelocytes, myelocytes, and promyelocytes. Manual differentials were performed on each specimen by 2 experienced medical technologists using the National Committee for Clinical Laboratory Standards (NCCLS) document H20-A. In addition to evaluating the IG parameter, St. Anthony's laboratory plans to report the automated IG count from the XE-2100, with