High-power and high-accuracy integrating sphere radiometer: design, characterization, and calibration.

Appl Opt

Instituto de Física Aplicada, Consejo Superior de Investigaciones Científicas, Madrid, Spain.

Published: January 2006

We describe the design, characterization, and calibration of a high-power and high-accuracy transfer standard for optical power measurements in fibers based on an integrating sphere radiometer working from -50 to +30 dBm. The integrating sphere radiometer has been calibrated in the spectral range 1250-1650 nm by use of an electrically calibrated pyroelectric radiometer and four tunable laser diodes. The total uncertainty obtained is less than +/-0.8% for these wavelength and power ranges.

Download full-text PDF

Source
http://dx.doi.org/10.1364/ao.45.000511DOI Listing

Publication Analysis

Top Keywords

integrating sphere
12
sphere radiometer
12
high-power high-accuracy
8
design characterization
8
characterization calibration
8
high-accuracy integrating
4
radiometer
4
radiometer design
4
calibration describe
4
describe design
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?