The enhancement of a dissolved chemical's Raman scattering by a liquid-core optical fiber (LCOF) geometry is absorption dependent. This dependence leads to a disruption of the usual linear correlation between chemical concentration and Raman peak area. To recover the linearity, we augmented a standard LCOF Raman spectroscopy system with spectrophotometric capabilities, permitting sequential measurements of Raman and absorption spectra within the LCOF. Measurements of samples with identical Raman-scatterer concentrations but different absorption coefficients are described. Using the absorption values, we reduced variations in the measured Raman intensities from 60% to less than 1%. This correction method should be important for LCOF-based Raman spectroscopy of sample sets with variable absorption coefficients, such as urine and blood serum from multiple patients.

Download full-text PDF

Source
http://dx.doi.org/10.1364/ao.45.000489DOI Listing

Publication Analysis

Top Keywords

correction method
8
liquid-core optical
8
optical fiber
8
raman spectroscopy
8
absorption coefficients
8
raman
6
absorption
5
method absorption-dependent
4
absorption-dependent signal
4
signal enhancement
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!