Purpose: To evaluate the relationship between the primary breast volume and dose received by the ipsilateral lung, heart (for left-breast cancers), and contralateral breast during primary breast irradiation using intensity-modulated radiation therapy (IMRT).
Methods And Materials: Sixty-five patients with breast carcinoma were treated using 6-MV photons with IMRT technique using the Eclipse Planning System following breast conserving surgery. All patients had a treatment planning CT scan. The primary breast, ipsilateral lung, and heart were contoured on the axial CT slices. The primary breast volume was calculated using the Eclipse Planning System. The mean ipsilateral lung and heart doses were obtained from the dose-volume histogram. The contralateral breast dose was measured using paired thermoluminescent dosimeters (TLDs) placed on the patient's contralateral breast, 4 cm from the center of the medial border of the primary breast irradiation field.
Results: The mean dose delivered with photons to the primary breast for all patients was 49.97 Gy. The mean volume of the primary irradiated breast was 1167.9 cc. As a percentage, the mean ipsilateral lung, heart, and contralateral breast doses were 11.2%, 6.1%, and 7.2%, respectively. The primary breast volume positively correlated with the contralateral breast dose (P < 0.0005). There was no significant correlation between the breast volume and the ipsilateral lung or heart dose (P = 0.463 and 0.943, respectively).
Conclusion: This study suggests that the primary breast size significantly affects the scatter dose to the contralateral breast but not the ipsilateral lung or heart dose when using IMRT for breast irradiation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/01.coc.0000198743.80991.15 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!