Objectives: To describe a strategy for revising the malfunctioning InterStim device and to provide an algorithm for evaluation and management.

Methods: We retrospectively reviewed 82 patients who had undergone InterStim placement. Ten patients (eight women and two men) experienced complications and subsequently underwent revision of their device between October 2001 and October 2003. Five patients had originally received a permanent implant after a successful percutaneous test stimulation trial, and five had undergone a test stimulation using the tined lead. Indications for revision included gradual onset of recurrent voiding dysfunction (n = 2), lead migration (n = 5), generator malfunction (n = 1), generator site pain and infection (n = 1), and genital/rectal pain with stimulation (sensory discomfort; n = 1).

Results: Of the 10 patients who underwent revision, 7 experienced complete resolution of their problem. Eight patients had lead site changes and two had generator replacements. No intraoperative or postoperative complications occurred in the revision cases.

Conclusions: In our experience, 70% of patients who undergo revision of the malfunctioning InterStim can expect success. In this study, no difference in success appeared to be related to the original cause of malfunction. In the management of malfunctioning sacral neuromodulators, we recommend an attempt at revision before permanent explantation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.urology.2005.08.055DOI Listing

Publication Analysis

Top Keywords

management malfunctioning
8
malfunctioning sacral
8
malfunctioning interstim
8
underwent revision
8
test stimulation
8
patients
6
revision
6
evaluation management
4
malfunctioning
4
sacral neuromodulator
4

Similar Publications

Purpose: The aim of this study is to examine the characteristics of intraoperative nursing near-miss events in interventional operating rooms, systematically identify and analyze associated risks, and propose effective mitigation strategies.

Patients And Methods: A retrospective study was conducted using a specially designed survey focused on nursing near-miss events in Interventional operating rooms. Records of intraoperative near-miss events voluntarily reported by medical and nursing staff between January 2023 and March 2024 were analyzed.

View Article and Find Full Text PDF

Deployable electronics with enhanced fatigue resistance for crumpling and tension.

Sci Adv

January 2025

Multiscale Bio-inspired Technology Lab, Department of Mechanical Engineering, Ajou University, 206 World cup-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do 16499, South Korea.

Highly packable and deployable electronics offer a variety of advantages in electronics and robotics by facilitating spatial efficiency. These electronics must endure extreme folding during packaging and tension to maintain a rigid structure in the deployment state. Here, we present foldable and robustly deployable electronics inspired by Plantago, characterized by their tolerance to folding and tension due to integration of tough veins within thin leaf.

View Article and Find Full Text PDF

Background: Public healthcare practices, particularly disease prevention, screening, diagnosis, treatment, and rehabilitation of patients, heavily rely on the availability and functionality of medical equipment. The absence of sufficient medical equipment and the malfunctioning of existing equipment impede the ability to provide effective healthcare services and directly affect patient rehabilitation, while the challenges related to medical equipment utilization are huge, especially in countries with limited resources such as Ethiopia. Therefore, this study aimed to assess the availability, functionality, and barriers associated with the use of medical equipment at public comprehensive specialized hospitals in Amhara Regional State, Ethiopia.

View Article and Find Full Text PDF

Objectives: We aimed to determine the compliance, safety, and acceptance of colon capsule endoscopy (CCE) and small bowel capsule endoscopy (SBCE) in an out-of-clinic setting remotely supported by medical staff.

Methods: We enrolled 30 examinees (24 with CCE and six with SBCE) who had not undergone CE at six gastroenterological centers. All examinees were provided with instructions on equipment and bowel preparations.

View Article and Find Full Text PDF

Unlabelled: Corrected transposition of the great arteries (c-TGA) is a congenital abnormality characterized by atrioventricular (AV) and ventriculoarterial discordance, associated with early and late-onset conduction disturbances. We report the case of a c-TGA patient affected by congenital complete AV block, with right ventricular lead malfunction with prolonged dwell time and two abandoned leads. He underwent leadless VDD pacemaker implantation at an unusual site (i.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!