Purpose: The aim of the current study was to show if contrast-to-noise ratio (CNR) could be improved without loss of diagnostic accuracy if a phase-sensitive inversion recovery (PSIR) single-shot TrueFISP sequence is used at 3.0 T instead of 1.5 T.

Material And Methods: Ten patients with myocardial infarction were examined on a 1.5 T magnetic resonance (MR) system (Avanto, Siemens Medical Systems) and at a 3.0 T MR system. Imaging delayed contrast enhancement was started 10 minutes after application of contrast material. A phase-sensitive inversion recovery (PSIR) single-shot TrueFISP sequence was used at 1.5 and 3.0 T and compared with a segmented IR turboFLASH sequence at 1.5 T, which served as the reference method. Infarct volumes and CNR of infarction and normal myocardium were compared with the reference method.

Results: The PSIR Single-Shot TrueFISP technique allows for imaging nine slices during a single breathhold without adaptation of the inversion time. The mean value of CNR between infarction and normal myocardium was 5.9 at 1.5 T and 12.2 at 3.0 T (magnitude images). The CNR mean value of the reference method was 8.4. The CNR mean value at 3.0 T was significantly (P = 0.03) higher than the mean value of the reference method. The correlation coefficients of the infarct volumes, determined with the PSIR single-shot TrueFISP technique at 1.5 T and at 3.0 T and compared with the reference method, were r = 0.96 (P = 0.001) and r = 0.99 (P = 0.0001).

Conclusion: The use of PSIR single-shot TrueFISP at 3.0 T allows for accurate detection and assessment of myocardial infarction. CNR is significantly higher at 3.0 T compared with 1.5 T. The PSIR single-shot technique at 3.0 T provides a higher CNR than the segmented reference technique at 1.5 T.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.rli.0000195843.97582.f4DOI Listing

Publication Analysis

Top Keywords

psir single-shot
28
single-shot truefisp
24
reference method
16
phase-sensitive inversion
12
inversion recovery
12
recovery psir
12
myocardial infarction
12
assessment myocardial
8
truefisp sequence
8
infarct volumes
8

Similar Publications

Purpose: Wideband phase-sensitive inversion recovery (PSIR) late gadolinium enhancement (LGE) enables myocardial scar imaging in implantable cardioverter defibrillators (ICD) patients, mitigating hyperintensity artifacts. To address subendocardial scar visibility challenges, a 2D breath-hold single-shot electrocardiography-triggered black-blood (BB) LGE sequence was integrated with wideband imaging, enhancing scar-blood contrast.

Methods: Wideband BB, with increased bandwidth in the inversion pulse (0.

View Article and Find Full Text PDF

Background: This study compares three-dimensional (3D) high-resolution (HR) late gadolinium enhancement (LGE; 3D HR-LGE) imaging using a respiratory navigated, electrocardiographically-gated inversion recovery gradient echo sequence with conventional LGE imaging using a single-shot phase-sensitive inversion recovery (PSIR) balanced steady-state free precession (bSSFP; PSIR-bSSFP) sequence for routine clinical use in the pediatric population.

Methods: Pediatric patients (0-18 years) who underwent clinical cardiovascular magnetic resonance (CMR) with both 3D HR-LGE and single-shot PSIR-bSSFP LGE between January 2018 and June 2020 were included. Image quality (0-4) and detection of LGE in the left ventricle (LV) (per 17 segments), in the right ventricle (RV) (per 3 segments), as endocardial fibroelastosis (EFE), at the hinge points, and at the papillary muscles was analyzed by two blinded readers for each sequence.

View Article and Find Full Text PDF

Myocardial motion-corrected phase-sensitive inversion recovery late gadolinium enhancement in free breathing paediatric patients: a comparison with single-shot coherent gradient echo ("TrueFISP") phase-sensitive inversion recovery.

Clin Radiol

June 2021

Department of Radiology, Key Laboratory of Obstetric & Gynecologic and Paediatric Diseases and Birth Defects of Ministry of Education, West China Second University Hospital, Sichuan University, 20# South Renmin Road, Chengdu, Sichuan, 610041, China. Electronic address:

Aim: To investigate the value of motion-corrected (MOCO) phase-sensitive inversion recovery (PSIR) late gadolinium enhancement (LGE) compared with single-shot balanced steady-state gradient echo ("TrueFISP", Siemens) PSIR in free breathing paediatric patients.

Materials And Methods: In this retrospective study, 238 paediatric patients underwent clinical contrast-enhanced cardiovascular magnetic resonance imaging (CMRI). Both the single-shot TrueFISP PSIR and MOCO PSIR sequences were performed on each child.

View Article and Find Full Text PDF

Objective: To study the application value of motion-correction phase sensitive inversion recovery (MOCO-PSIR) to evaluate myocardial fibrosis in the patients with heart failure caused by dilated cardiomyopathy (DCM).

Methods: A prospective study included 60 patients who underwent cardiac MRI enhanced scan from June 2017 to November 2018, including 38 patients who were clinically diagnosed with DCM and 22 patients in the normal control group. All patients were scanned with three late gadolinium enhancement (LGE) sequences: segmented-PSIR, single-shot-PSIR, MOCO-PSIR at the same time.

View Article and Find Full Text PDF

Background: Segmented phase-sensitive inversion recovery (PSIR) cardiovascular magnetic resonance (CMR) sequences are reference standard for non-invasive evaluation of myocardial fibrosis using late gadolinium enhancement (LGE). Several multi-slice LGE sequences have been introduced for faster acquisition in patients with arrhythmia and insufficient breathhold capability. The aim of this study was to assess the accuracy of several multi-slice LGE sequences to detect and quantify myocardial fibrosis in patients with ischemic and non-ischemic myocardial disease.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!