Ten fresh mandibles from adult sheep were stripped of all soft tissues and sectioned in the midline. We did sagittal split osteotomies and 5 mm advancement on all the 20 hemimandibles. Ten hemimandibles were fixed with three 2.0 mm x 13 mm titanium bicortical screws, and the other 10 were fixed with three 2.0 mm x 13 mm poly-l-lactic acid/polyglycolic acid (PLLA/PGA) bicortical screws in an inverted L pattern. All the hemimandibles were then mounted in a servohydraulic testing unit and tested to permanent deformation. Maximum forces that the mandibles resisted before breaking, maximum displacements, and the displacement values under 20, 60, 120, and 150 N were compared using the Mann-Whitney U-test. There were no significant differences in stability between the bones fixed with titanium and those fixed with resorbable screws.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.bjoms.2005.11.020DOI Listing

Publication Analysis

Top Keywords

bicortical screws
12
sagittal split
8
fixed resorbable
8
titanium bicortical
8
fixed three
8
fixed
5
mechanical stability
4
stability sagittal
4
split ramus
4
ramus osteotomy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!