Background: Defibrillation testing (DT) at the end of the implantation of cardiac resynchronization pacemaker with a defibrillator (CRT-D) exposes heart failure (HF) patients to increased procedural risks. However, until now, delayed DT has not been assessed as a possible option in HF patients implanted with CRT-D.
Objective: Aim of the present study is to assess safety and feasibility of delayed DT in HF patients treated with CRT-D.
Material And Methods: Two hundred and eleven consecutive patients (mean age: 65 years, mean NYHA class 3.0, mean EF: 29.3%) underwent CRT-D implantation from October 1999 to December 2004. In the first 17 patients, DT was performed at the end of CRT-D implantation. In the other 194 consecutive patients, DT was performed at 2 months after CRT-D implantation. Outcome of DT, as well as "acute" LV lead dislodgment rate were evaluated in the latter group of 194 patients undergoing a delayed DT. Also, ICD function was assessed through device telemetry analysis at 2 months.
Results: At delayed DT, first shock was effective in 187 of 194 patients (96%), ineffective VF interruption at maximum energy occurred only in one patient (0.5%), and acute LV lead dislodgment was 1%. No ICD therapy failure occurred in the 2-month untested period.
Conclusion: DT performed 2 months after CRT-D implantation is safe and feasible; this is possibly related to the improvement of clinical conditions and hemodynamic status as well as greater lead stability 2 months after CRT-D.