We compared a new serum immunoassay for quantitation of serum free light chains (FLC) with the conventional tests for clonal immunoglobulin production: bone marrow immunohistochemistry, serum immunofixation electrophoresis, and urine immunofixation electrophoresis. Serum samples from 169 patients with AL amyloidosis and 20 controls were examined. Elevated levels of kappa-FLC and lambda-FLC were found in 94% and 93% of patients with the respective clonal disease. However, false positive elevations of kappa-FLC and lambda-FLC were found in 30% and 44% of patients with clonal disease of the other light chain subtype. We found that the FLC level was a reliable test for the diagnosis of clonal disease when the FLC kappa:lambda ratio was abnormal and was comparable to the conventional tests in patients with AL amyloidosis. After a histologic tissue diagnosis of amyloidosis, determining the type as AL amyloidosis relies on a panel of hematologic tests to determine light chain clonality and the exclusion other forms of amyloidosis.

Download full-text PDF

Source
http://dx.doi.org/10.1080/13506120500352339DOI Listing

Publication Analysis

Top Keywords

light chain
12
clonal disease
12
serum free
8
free light
8
conventional tests
8
immunofixation electrophoresis
8
patients amyloidosis
8
kappa-flc lambda-flc
8
amyloidosis
6
quantitative serum
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?