This paper reports on a novel continuous-flow hydrogenation reactor and its integration with a liquid handler to generate a fully automated high-throughput hydrogenation system for library synthesis. The reactor, named the H-Cube, combines endogenous hydrogen generation from the electrolysis of water with a continuous flow-through system. The system makes significant advances over current batch hydrogenation reactors in terms of safety, reaction validation efficiency, and rates of reaction. The hydrogenation process is described along with a detailed description of the device's main parts. The reduction of a series of functional groups, varying in difficulty up to 70 degrees C and 70 bar are also described. The paper concludes with the integration of the device into an automated liquid handler followed by the reduction of a nitro compound in a high throughput manner. The system is fully automated and can conduct 5 reactions in the time it takes to perform and workup one reaction manually on a standard batch reactor.

Download full-text PDF

Source
http://dx.doi.org/10.1021/cc050107oDOI Listing

Publication Analysis

Top Keywords

hydrogenation reactor
8
liquid handler
8
fully automated
8
hydrogenation
5
continuous-flow high
4
high pressure
4
pressure hydrogenation
4
reactor
4
reactor optimization
4
optimization high-throughput
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?