The velocity distribution of liquid flowing in a commercial micromixer has been determined directly by using pulsed-field gradient NMR. Velocity maps with a spatial resolution of 29 microm x 43 microm were obtained by combining standard imaging gradient units with a homebuilt rectangular surface coil matching the mixer geometry. The technique provides access to mixers and reactors of arbitrary shape regardless of optical transparency. Local heterogeneities in the signal intensity and the velocity pattern were found and serve to investigate the quality and functionality of a micromixer, revealing clogging and inhomogeneous flow distributions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1039/b510708c | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!