Purpose: The variable-pitch pulse oximeter is an important intraoperative patient monitor. Our ability to hear its auditory signal depends on its acoustical properties and our hearing. This study quantitatively describes the audio spectrum and sound pressure levels of the monitoring tones produced by five variable-pitch pulse oximeters.

Methods: We compared the Datex-Ohmeda Capnomac Ultima, Hewlett-Packard M1166A, Datex-Engstrom AS/3, Ohmeda Biox 3700, and Datex-Ohmeda 3800 oximeters. Three machines of each of the five models were assessed for sound pressure levels (using a precision sound level meter) and audio spectrum (using a hanning windowed fast Fourier trans-form of three beats at saturations of 99%, 90%, and 85%).

Results: The widest range of sound pressure levels was produced by the Hewlett-Packard M1166A (46.5 +/- 1.74 dB to 76.9 +/- 2.77 dB). The loudest model was the Datex-Engstrom AS/3 (89.2 +/- 5.36 dB). Three oximeters, when set to the lower ranges of their volume settings, were indistinguishable from background operating room noise. Each model produced sounds with different audio spectra. Although each model produced a fundamental tone with multiple harmonic overtones, the number of harmonics varied with each model; from three harmonic tones on the Hewlett-Packard M1166A, to 12 on the Ohmeda Biox 3700. There were variations between models, and individual machines of the same model with respect to the fundamental tone associated with a given saturation.

Conclusion: There is considerable variance in the sound pressure and audio spectrum of commercially-available pulse oximeters. Further studies are warranted in order to establish standards.

Download full-text PDF

Source
http://dx.doi.org/10.1007/BF03021524DOI Listing

Publication Analysis

Top Keywords

sound pressure
20
audio spectrum
16
pressure levels
16
hewlett-packard m1166a
12
spectrum sound
8
pulse oximeters
8
variable-pitch pulse
8
datex-engstrom as/3
8
ohmeda biox
8
biox 3700
8

Similar Publications

Telehealth is increasing popular as a treatment option for people with Parkinson disease (PD). The SpeechVive device is a wearable device that uses the Lombard effect to help patients speak more loudly, slowly, and clearly. This study sought to examine the effectiveness of the device to improve communication in people with PD, delivered over a telehealth modality as compared to in-person, using implementation science design.

View Article and Find Full Text PDF

Carbon Black Absorption Enhanced Fiber-Optic Photoacoustic Gas Sensing System with Ultrahigh Sensitivity.

Anal Chem

January 2025

School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, Liaoning 116024, China.

A highly sensitive trace gas sensing system based on carbon black absorption enhanced photoacoustic (PA) spectroscopy (PAS) is reported. A carbon black sheet and a fiber-optic cantilever microphone (FOCM) are integrated to form a fiber-optic cantilever spectrophone (FOCS). The gas concentration is obtained by measuring the acoustic wave amplitude generated by the carbon black sheet, which absorbs the laser passing through the interest gas.

View Article and Find Full Text PDF

Background: Bubble Continuous Positive Airway Pressure (bCPAP) has become a crucial intervention in neonatal care, particularly in resource-limited settings where access to advanced respiratory support is scarce. While bCPAP offers a cost-effective solution for neonatal respiratory distress, conducting clinical trials in these settings presents significant ethical challenges. Issues of justice, beneficence, and autonomy arise due to disparities in healthcare infrastructure, complexities surrounding informed consent, and the potential exploitation of vulnerable populations.

View Article and Find Full Text PDF

In this study, the effectiveness of three choline chloride (ChCl)-based deep eutectic solvents (DESs) formed using malonic acid (MalA), glycerol (Gly), and glucose (Glu) as hydrogen bond donors and two conventional solvents (50% methanol and 50% ethanol) for ultrasonic-assisted extraction (UAE) of antioxidant compounds from four herbs (chamomile, lemon balm, nettle, and spearmint) were estimated. The antioxidant capacity (AC) of the obtained herb extracts was determined by the modified 2,2'-azino-bis(3-ethylbenzothiazoline-6-sulfonic acid) (ABTS), 2,2-diphenyl-1-picrylhydrazyl (DPPH), and cupric reducing antioxidant capacity (CUPRAC) methods. Profiles of phenolic acids, flavonoid aglycones, and flavonoid glycosides in the green and conventional herb extracts were quantitatively analyzed using ultra-performance liquid chromatography (UPLC).

View Article and Find Full Text PDF

L., commonly known as the mastic tree or lentisk, is a woody Mediterranean plant revered for its ecological relevance as well as for its extensive ethnobotanical heritage. Historically, the fruits and the resin of have been widely utilized in traditional medicine, underscoring its important role in local healing practices.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!