Background And Purpose: Various laparoscopic devices have been described for suture anchoring during solidorgan parenchymal closure. Application of these devices expedites the closure of parenchymal defects and minimizes ischemia time. We compared different technologies as suture anchors for parenchymal closure.

Materials And Methods: A tensometer was used to determine the amount of tension necessary to dislodge each of five different clips from Vicryl suture alone or against two different substrates (fresh pig kidney and liver) with and without an intervening pledget. The clips investigated were the Lapra-Ty (Ethicon), Endoclip II (US Surgical), small Horizon Ligating Clips (Weck), Hem-o-lok Medium Polymer Clips (Week), and a novel Suture-clip (Applied Medical). ANOVA and two-sided Fisher's exact test provided statistical analysis.

Results: The force required to dislodge the Lapra-Ty clip from bare suture for both 0 and 1 Vicryl (7.0 N) was approximately fourfold the force required to dislodge the Endoclips or the 5-mm or 10-mm Hem-o-lok clips (p<0.01). When clips were applied to suture running through renal or liver parenchyma, the novel Suture-clip required the greatest tension to dislodge (P<0.01), followed by the Horizon and Lapra-Ty clips. There were no statistically significant differences in the tension required to dislodge a given clip from the two parenchymal substrates or in the presence or absence of a pledget.

Conclusions: In our experimental model, the Suture-clip, Lapra-Ty, and Horizon clips required significantly greater tension to dislodge than the Hem-o-lok and Endoclip clips. The addition of a pledget did not improve tension resistance.

Download full-text PDF

Source
http://dx.doi.org/10.1089/end.2005.19.1221DOI Listing

Publication Analysis

Top Keywords

suture anchors
8
force required
8
required dislodge
8
suture
5
clips
5
comparison holding
4
holding strength
4
strength suture
4
anchors hepatic
4
hepatic renal
4

Similar Publications

Background: Arthroscopic labral repair of the hip is successfully performed with increasing frequency using either knotless or knotted suture anchors, each with its own risks and benefits.

Purpose: To examine biomechanical and clinical outcomes for labral repair of the hip based on the use of knotted or knotless suture anchors.

Study Design: Systematic review; Level of evidence, 4.

View Article and Find Full Text PDF

Objective: To compare the 3-year outcomes of the modified minimally invasive Ponto surgery (m-MIPS) to both the original MIPS (o-MIPS) and linear incision technique with soft tissue preservation (LIT-TP) for inserting bone-anchored hearing implants (BAHIs).

Study Design: Prospective study with three patient groups: m-MIPS, o-MIPS, and LIT-TP.

Setting: Tertiary referral center.

View Article and Find Full Text PDF

The clinicopathologic conditions of the long head of the biceps tendon vary, encompassing tendinitis, peritendinous inflammation, hypertrophy, and partial or complete tears. These symptoms are typically linked with SLAP tears and instability of the long head of the biceps tendon, often resulting in partial displacement or complete dislocation. The choice between tenotomy and tenodesis depends on varied factors.

View Article and Find Full Text PDF

Anatomic Distal Biceps Tendon Repair With All-Suture Cortical Buttons.

Arthrosc Tech

December 2024

Department of Orthopaedic Surgery, University of California Irvine, Orange, California, U.S.A.

Acute, traumatic distal biceps tendon ruptures are a common injury in the middle-aged athletic male population, with direct anatomic surgical repair being the most effective technique to restore maximal strength. Multiple techniques for distal biceps tendon repair have been described, including single- or dual-incision approaches and tendon fixation with cortical buttons, interference screws, suture anchors, and transosseous sutures. In this Technical Note, we demonstrate an anatomic distal biceps tendon repair technique with a single-incision approach using 2 all-suture cortical buttons.

View Article and Find Full Text PDF

Remplissage is a procedure that decreases the rate of recurrent instability after arthroscopic stabilization in patients with an engaging Hill-Sachs lesion. This technique involves capsulotenodesis of the infraspinatus tendon and posterior capsule into the Hill-Sachs lesion using 2 knotless anchors with suture passage through the infraspinatus tendon guided by a percutaneous needle. Previously described techniques use knots or anchor placement through the infraspinatus, which can be challenging to control and irreversible if tendon penetration occurs in an undesirable location.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!