This paper describes Monte Carlo simulation of intrinsic count rate performance of a scintillation gamma camera. MCNP Monte Carlo code was employed to calculate pulse height spectrum and detector efficiency. A custom code written in Fortran language was then developed to simulate, by Monte Carlo method, the distortion in pulse height spectrum due to the pile-up effect for paralyzable and nonparalyzable systems. The results of the simulations, compared with the experimental measurement of count rate performance, showed a good agreement between the two different approaches.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.apradiso.2005.08.021 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!