Background: Dobutamine stress MR (DSMR) is increasingly used in the clinical routine. Due to the limitations inherent to MR, i.e. the distance of the personnel to the patient during imaging, it is necessary to set up an emergency training to rescue a patient from the scanner room in the shortest possible time.

Methods: We placed a volunteer in the 1.5 T scanner. Monitoring devices are attached in the same way as a patient would be prepared. A team of four performed the drill, everybody assigned to a certain task and position of the rescue procedure. In consecutive order, scanning was stopped, the MR table was released and the team entered the scanner room. The monitoring equipment was removed from the volunteer and he was slid onto a stretcher and moved out of the scanner room.

Results: The drill was performed 30 times, 3 teams were involved. The minimum time required was 32 s, maximum was 52 s. The amount of practice runs significantly reduced the rescue time. Team 1 with 15 runs had an average time of 39+/-6 s vs. team 2 with 9 runs (45+/-3 s, p<0.02) or team 3 with 6 runs (46+/-3 s, p<0.01). Team 1 practiced on three separate days each with five practice runs and consecutively improved from day one (43+/-5 s) to day three (35+/-2 s, p<0.02).

Conclusion: A significant reduction in rescue time can be achieved with repetitive exercise. We recommend repetitive training with various scenarios, volunteers and personnel involved in the DSMR testing.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10554-005-9028-9DOI Listing

Publication Analysis

Top Keywords

scanner room
8
team runs
8
scanner
5
code blue
4
blue suite--a
4
suite--a drill
4
rescue
4
drill rescue
4
rescue patients
4
patients cardiac
4

Similar Publications

Reproducibility of quantitative myocardial perfusion and coronary flow capacity by positron emission tomography: 3D digital silicon photomultiplier solid state vs. legacy 2D analogue systems for clinical practice and trials.

Eur Heart J Imaging Methods Pract

July 2024

Weatherhead PET Center for Preventing and Reversing Atherosclerosis, Division of Cardiology, Department of Medicine, McGovern Medical School, University of Texas Health Science Center at Houston, Memorial Hermann Hospital, 6431 Fannin St., Room MSB 4.256 Houston, TX 77030, USA.

Aims: Quantitative rest-stress myocardial perfusion in millilitres per minute per gram among multiple 2D and 3D positron emission tomography-computed tomography (PET-CT) scanners is essential for personalized cardiac management and clinical trials. Accordingly, this study reports the accuracy and precision of quantitative rest-stress millilitres per minute per gram and coronary flow capacity among 2D and two different digital 3D silicon photomultiplier (SiPM) PET-CT scanners for quantifying the severity of coronary pathophysiology for clinical trials or guiding interventions vs. medical treatment.

View Article and Find Full Text PDF

This paper explores the influence of various camera settings on the quality of 3D reconstructions, particularly in indoor crime scene investigations. Utilizing Neural Radiance Fields (NeRF) and Gaussian Splatting for 3D reconstruction, we analyzed the impact of ISO, shutter speed, and aperture settings on the quality of the resulting 3D reconstructions. By conducting controlled experiments in a meeting room setup, we identified optimal settings that minimize noise and artifacts while maximizing detail and brightness.

View Article and Find Full Text PDF

Background: Anxiety-provoking healthcare procedures require specific child-friendly approaches. Magnetic resonance imaging (MRI) can cause anxiety for children and general anesthesia (GA) is often used. We developed and tested a multi-faceted child-friendly concept, Children Centered Care, for MRI of children without GA.

View Article and Find Full Text PDF

Our autobiographical experiences typically occur within the context of familiar spatial locations. When we encode these experiences into memory, we can use our spatial map of the world to help organize these memories and later retrieve their episodic details. However, it is still not well understood what psychological and neural factors make spatial contexts an effective scaffold for storing and accessing memories.

View Article and Find Full Text PDF

Rationale And Objectives: Multi-parametric MRI (mpMRI) studies of the body are routinely acquired in clinical practice. However, a standardized naming convention for MRI protocols and series does not exist currently. Conflicts in the series descriptions present in the DICOM headers arise due to myriad MRI scanners from various manufacturers used for imaging, wide variations in imaging practices across institutions, and technologist preferences.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!