Background: We have demonstrated previously that triphasic waveform shocks were superior to biphasic waveform shocks for transthoracic defibrillation. Our purpose was to compare the efficacy and safety of quadriphasic versus triphasic shocks for transthoracic defibrillation in a porcine model.

Methods: Sixteen adult swine (19-25 kg, mean: 21.5 kg) were deeply anesthetized and intubated. To simulate impedance of the human chest, fixed electrical resistors (25 or 50 ohms) was placed in series with the defibrillator and the chest of each pig. After 30 s of electrically induced VF, each pig received transthoracic shocks, using either a truncated exponential triphasic waveform (5 ms positive pulse duration, 5 ms negative pulse duration and 5 ms positive pulse duration, total waveform duration 15 ms) or a quadriphasic waveform (5/5/5/5 ms, total waveform duration 20 ms). Each pig received transthoracic triphasic and quadriphasic shocks at three selected energy levels (50, 100 and 150 J) in random sequence. Four shocks were delivered at each energy level to construct an energy versus % success curve. Success was defined as VF termination at 5 s after shock. The total shocks were divided into three groups based on the delivered energy actually delivered to the animal: <40, 40-65 and >65 J. Delivered energy = (animal impedance/total impedance) times selected energy of the shock.

Results: For high-impedance animals (86-102 ohms), quadriphasic waveform shocks achieved significantly higher percent shock success than triphasic shocks for the termination of VF at the energy levels of >65 J actually delivered (quadriphasic 72.7+/-12.2%, triphasic 38.9+/-7.7%, p<0.02). No differences in the shock success between quadriphasic and triphasic waveforms were found for other two energy levels. There were no differences in ventricular tachycardia or asystole after shocks between quadriphasic and triphasic waveforms.

Conclusion: In this porcine model, 20 ms (5/5/5/5) quadriphasic shocks were superior to 15 ms (5/5/5) triphasic shocks for transthoracic defibrillation in animals with impedances that simulated high impedance in humans.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.resuscitation.2005.05.025DOI Listing

Publication Analysis

Top Keywords

transthoracic defibrillation
12
waveform shocks
12
pulse duration
12
delivered energy
12
shocks
9
triphasic waveform
8
shocks transthoracic
8
triphasic shocks
8
pig received
8
received transthoracic
8

Similar Publications

We describe, to our knowledge, the first use in Dubai of extracorporeal life support (ECLS) in a patient who suffered intraoperative cardiac arrest due to presumed cardiac channelopathy. A 40-year-old patient presented for open myomectomy surgery. She had no other medical problems apart from obesity.

View Article and Find Full Text PDF

Cardiac implantable electronic devices and their integrated thoracic impedance sensors have been used to detect sleep apnea for over a decade now. Despite their usage in daily clinical practice, there are only limited data on their diagnostic accuracy. AIRLESS and UPGRADE were prospective investigator-driven trials meant to validate the AP scan (Boston Scientific, Marlborough, MA, USA) in heart failure cohorts.

View Article and Find Full Text PDF

Background: Anomalous origin of the left coronary artery (LCA) from the pulmonary artery (PA) (ALCAPA) is a rare congenital abnormality. We present a case of an ALCAPA in a 25-year-old man.

Case Summary: A 25-year-old male with no past medical history was admitted to our intensive cardiac care unit after sudden cardiac arrest due to ventricular fibrillation and suspected acute coronary syndrome.

View Article and Find Full Text PDF

Background: Hypertrophic cardiomyopathy (HCM) needs careful differentiation from other cardiomyopathies. Current guidelines recommend genetic testing, but genetic data on differential diagnoses and their relation with clinical outcomes in HCM are still lacking. This study aimed to investigate the prevalence of genetic variants and the proportion of other cardiomyopathies in patients with suspected HCM in Korea and compare the outcomes of HCM according to the presence of sarcomere gene mutation.

View Article and Find Full Text PDF
Article Synopsis
  • ARVC is a rare genetic condition that increases the risk of serious heart problems and requires effective risk assessment to prevent severe cardiac events.
  • A study involving 83 ARVC patients showed that those experiencing major adverse cardiac events (MACE) had notably lower right ventricular global longitudinal strain (RV GLS) and right ventricular free wall longitudinal strain (RVFWLS) compared to those without MACE.
  • The findings suggest that measuring RVFWLS can be a better predictor of MACE than traditional echocardiographic methods for assessing right ventricular function.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!