Aims: To compare the values measured for retinal macular thickness with the first and last generations of the optical coherence tomograph (OCT1 and Stratus OCT, Zeiss, Humphrey Division).
Methods: This was a cohort study. 59 eyes were examined: 17 had a normal macula and 42 had a diabetic macular oedema. In each eye, mean retinal thickness (RT) was measured automatically in the nine macular Early Treatment Diabetic Retinopathy Study areas and at the foveal centre, using OCT1 and Stratus OCT. The paired mean RT values for each area and the type and proportion of artefacts were compared.
Results: Of the 590 automatic measurements, 505 had no artefact, either with OCT1 or Stratus OCT. The mean difference between the OCT1 and Stratus OCT measurements was 25 (SD 26.2) microm (p<0.0001). With Stratus OCT, RT values were significantly higher, by 8.1% (7.8%), than with OCT1. Artefacts were only observed in cases of diabetic macular oedema and were significantly more frequent with OCT1 than Stratus OCT (10.5% versus 4.4, p<0.0001).
Conclusion: The macular retinal thickness values measured with Stratus OCT were significantly higher than those measured with OCT1. Stratus OCT has the advantage of producing fewer artefacts than OCT1 in pathological cases.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1772971 | PMC |
http://dx.doi.org/10.1136/bjo.2005.069815 | DOI Listing |
Arq Bras Oftalmol
May 2007
Divisão Oftalmológica, Hospital das Clínicas, Faculdade de Medicina, Universidade de São Paulo, São Paulo, SP, Brasil.
Purpose: To compare the abilitiy of scanning laser polarimetry (GDx), optical coherence tomography (OCT) 1 and Stratus- optical coherence tomography to discriminate between healthy eyes and eyes with band atrophy of the optic nerve and temporal hemianopsia.
Methods: Twenty-three eyes with band atrophy of the optic nerve and 23 eyes from healthy subjects were included in this observational prospective case-control study. All eyes underwent retinal nerve fiber layer (RNFL) thickness analysis using GDx, optical coherence tomography-1 and Stratus-optical coherence tomography.
Br J Ophthalmol
December 2005
Service d'Ophtalmologie, Hôpital Lariboisière 2, rue Ambroise Paré, 75475 Paris Cedex 10, France.
Aims: To compare the values measured for retinal macular thickness with the first and last generations of the optical coherence tomograph (OCT1 and Stratus OCT, Zeiss, Humphrey Division).
Methods: This was a cohort study. 59 eyes were examined: 17 had a normal macula and 42 had a diabetic macular oedema.