Multiinstrument study of the electrophoretic mobility of quartz.

J Colloid Interface Sci

Department of Electrochemistry, Technical University of Lublin, Lublin, Nadbystrzycka 38A, 20618, Poland.

Published: June 2002

The electrophoretic mobility of quartz in 0.01 mol dm(-3) NaCl and NaNO3 over the pH range 2-8 was studied using five different commercial instruments. The mobilities over the pH range 4-8 were relatively consistent, but the mobilities over the pH range 2-4 and the position of the isoelectric point IEP varied from one instrument to another. This result suggests that the discrepancies in the apparent IEP of quartz (and other silicas) reported in the literature are due to the instrument artifacts.

Download full-text PDF

Source
http://dx.doi.org/10.1006/jcis.2002.8330DOI Listing