Background: The objectives of this pilot study were to (1) determine whether bipolar radiofrequency plasma-mediated ablation (PMA) can efficiently remove nasal septal cartilage and (2) calculate the ablation rate as a function of device power, probe force, and translation velocity using ex vivo porcine tissue.

Methods: Specimens were secured to a linear translation stage and were subjected to varying translation velocities (4, 7, and 12 cm/s), probe forces (140, 200, and 225 g), and bipolar radiofrequency generator (Coblator ENTec power settings, 38-58, 77-115, and 129-193 Watts root mean squared. Specimen mass loss and depth of ablation were measured using an analytic balance and dissection microscope, respectively.

Results: Visual and microscopic inspection revealed little char. Mass loss increased with decreasing translation velocity and increasing generator setting. Increasing probe force also influenced mass loss and increased crater depth. Depth of ablation did not correlate with translation velocity or generator setting.

Conclusion: PMA effectively ablates nasal septal cartilage and may be able to reduce or contour cartilaginous deformities and framework structures in the head and neck.

Download full-text PDF

Source

Publication Analysis

Top Keywords

bipolar radiofrequency
12
nasal septal
12
septal cartilage
12
translation velocity
12
mass loss
12
radiofrequency plasma-mediated
8
plasma-mediated ablation
8
probe force
8
depth ablation
8
loss increased
8

Similar Publications

The surgical management of hepatic metastases from colorectal cancer may range from segmental resections to major or extended hepatectomies. The aim is to achieve complete removal of metastatic lesions while preserving adequate liver function. We present the case of a 42-year-old male patient with a history of glucose intolerance who presented with altered bowel movements and abdominal pain.

View Article and Find Full Text PDF

Background: Epicardial ventricular tachycardia (VT) ablation is an established approach in patients with epicardial arrhythmogenic foci and is most commonly performed via percutaneous access. An alternative approach is via video-assisted thoracoscopic surgery (VATS), although reports of this technique are limited to the use of catheter-based technologies for radiofrequency ablation delivery.

Case Summary: A 55-year-old man with non-ischaemic cardiomyopathy presented with recurrent VT despite medical therapy.

View Article and Find Full Text PDF

Background: Psoriasis is a common chronic systemic disease affecting the skin, nails, and joints. Nails are commonly associated with a greater severity of the disease. Radiofrequency (RF) is a nonionizing radiation that provides energy originating from electric current to generate heat inside the dermis with anti-inflammatory effects.

View Article and Find Full Text PDF
Article Synopsis
  • The study examines the effectiveness of different energy sources used in the Cox-maze IV surgical procedure for treating atrial fibrillation (AF), specifically comparing bipolar radio-frequency (BRF), cryoenergy (Cryo), and a combination of both.
  • Results indicate that patients who underwent cryoablation had significantly lower rates of AF recurrence after both 6 months (6.73% for Cryo) and 4 years (6.14% for Cryo) compared to those treated with BRF and the combined method.
  • The findings suggest that cryoablation may lead to better long-term sinus rhythm restoration in patients undergoing AF ablation compared to bipolar radiofrequency ablation alone.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!