A methodology for 3D image reconstruction from retrospectively gated cone-beam CT projection data has been developed. A mobile x-ray cone-beam device consisting of an isocentric C-arm equipped with a flat panel detector was used to image a moving phantom. Frames for reconstruction were retrospectively selected from complete datasets based on the known rotation of the C-arm and a signal from a respiratory monitor. Different sizes of gating windows were tested. A numerical criterion for blur on the reconstructed image was suggested. The criterion is based on minimization of an Ising energy function, similar to approaches used in image segmentation or restoration. It is shown that this criterion can be used for the determination of the optimal gating window size. Images reconstructed from the retrospectively gated projection sequences using the optimal gating window data showed a significant improvement compared to images reconstructed from the complete projection datasets.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/50/22/004DOI Listing

Publication Analysis

Top Keywords

isocentric c-arm
8
reconstruction retrospectively
8
retrospectively gated
8
optimal gating
8
gating window
8
images reconstructed
8
respiratory correlated
4
correlated cone-beam
4
cone-beam computed
4
computed tomography
4

Similar Publications

Article Synopsis
  • The AAPM Task Group Report 135.B addresses updates and new technology features in an existing radiosurgery platform that weren't thoroughly covered in previous reports.
  • This task group combines foundational guidelines for new processes with educational resources aimed at helping medical physicists who are new to this technology.
  • Due to the limited user base of this technology compared to others, the report emphasizes educational components and provides prescriptive recommendations as some new features lack sufficient long-term data for comprehensive risk analysis.
View Article and Find Full Text PDF

Background: GammaKnife (GK) and CyberKnife (CK) have been the mainstay stereotactic radiosurgery (SRS) solution for multiple brain metastases (MBM) for several years. Recent technological advancement has seen an increase in single-isocentre C-arm linac-based SRS. This systematic review focuses on dosimetric and geometric insights into contemporary MBM SRS and thereby establish if linac-based SRS has matured to match the mainstay SRS delivery systems.

View Article and Find Full Text PDF

Isocentric fixed angle irradiation-based DRR: a novel approach to enhance x-ray and CT image registration.

Phys Med Biol

May 2024

Beijing Institute of Heart, Lung and Blood Vessel Diseases, Beijing Anzhen Hospital, Capital Medical University, Beijing, People's Republic of China.

Digitally reconstructed radiography (DRR) plays an important role in the registration of intraoperative x-ray and preoperative CT images. However, existing DRR algorithms often neglect the critical isocentric fixed angle irradiation (IFAI) principle in C-arm imaging, resulting in inaccurate simulation of x-ray images. This limitation degrades registration algorithms relying on DRR image libraries or employing DRR images (DRRs) to train neural network models.

View Article and Find Full Text PDF

Purpose: The purpose of this study is to evaluate the dosimetric and treatment delivery characteristics of volumetric modulated arc therapy technique (VMAT)-based craniospinal axis irradiation (CSI) between ring gantry Halcyon (HAL) and C-arm based Novalis Tx (NTx) linear accelerator. Set-up margin and treatment delivery time for both machines were also taken into account.

Materials And Methods: Fifteen patients, 4 females and 11 males treated between March 2019 and February 2022 within the age group 4-56 years simulated in the supine position and were planned for multiple isocentre VMAT technique in ring gantry Halcyon and C-Arm Novalis linear accelerator for 6FFF and 6 MV flatten beam energy.

View Article and Find Full Text PDF

3D C-arm navigated acromioclavicular joint stabilization.

Arch Orthop Trauma Surg

February 2024

Department of Trauma Hand and Reconstructive Surgery, Ulm University, Albert-Einstein-Allee 23, 89081, Ulm, Germany.

Introduction: Surgical treatment options for acromioclavicular joint separations are varied. Frequently, suspension devices (SD) are inserted for stabilization under arthroscopic view. This study investigates the feasibility and accuracy of three-dimensional (3D) digital-volume-tomography (DVT) C-arm navigated implantation with regard to the general trend toward increasingly minimally invasive procedures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!