Background: Five to 15% of the population have allergy to nickel, chromium, or molybdenum, which is a potential cause for in-stent restenosis. The Titan stent is made of stainless steel and is coated with titanium-nitride oxide (TiNOX), which completely prevents the discharge of metal elements. We performed a real-life multicenter registry to assess the short- and long-term characteristics of the Titan stent.
Methods And Results: A total of 103 Titan stents was implanted in 100 patients. Patients were 61.4+/-12.6 years old (81 men). Risk factors included hypercholesterolemia (63%), hypertension (53%), diabetes mellitus (DM; 35%), and current smoking (23%). Indications for PCI (percutaneous coronary intervention) were acute coronary syndromes (ACS) in 68% [acute ST elevation myocardial infarction (MI) in 8%], stable AP (angina pectoris) in 25%, and silent ischemia in 7% of the patients. Fifty-two percent of the treated lesions were of Type B2 or C. Lesion length was 14.3+/-2.9 mm and stent diameter was 3.06+/-0.36 mm. Indications for stenting were prevention of restenosis in 66%, residual stenosis in 33%, dissection in 13%, acute MI in 13%, and in-stent restenosis in 7% of the patients. Procedural success was 100%, with no complications. At 30 days, there were no major adverse cardiac events (MACE), including death, MI, and revascularization. At 180 days, only three patients had TVR (target vessel revascularization); two had TLR (target lesion revascularization) (one PCI and one CABG [coronary artery bypass grafting]), and one patient had a new narrowing proximal to the stent and underwent CABG due to multivessel disease.
Conclusions: The Titan stent has a remarkable safety profile in high-risk patients and complex coronary lesions and excellent short- and long-term outcome with a very low clinical TLR rate.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.carrev.2005.04.004 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!