Objective: The purpose of this cadaveric study was to compare the mechanical behavior of a locked compression plate, which uses threaded screw heads to create a fixed angle construct, with a dynamic compression plate construct in a cadaver radius model.

Design: Mechanical study with cyclic testing and high-speed optical motion analysis.

Setting: Biomechanics laboratory at an academic institution.

Patients/participants: Eighteen pairs of fresh-frozen human cadaver radii were divided into 3 groups of 6 to be tested as a group in each of the following force applications: anteroposterior (AP) bending, mediolateral bending, or torsion.

Intervention: Each bone was osteotomized leaving a 5-mm fracture gap and then fixed with a plate. For each pair, 1 radius received a standard plate (limited-contact dynamic compression plates; LC-DCP), the contralateral radius was fixed with a locking compression plate (LCP), and specimens underwent cyclic loading. Normalized stiffness, average energy absorbed, and Newton-cycles to failure were calculated. In addition, a 3-dimensional, high-speed, infrared motion analysis system was used to evaluate motion at the fracture site.

Main Outcome Measurements: Construct stiffness, fracture site motion, cycles to failure, and energy absorption. Repeated measures ANOVA were used to detect differences between groups with time.

Results: In the torsion group, LCP specimens failed at 60% greater Newton-cycles than the LC-DCP (1473 vs. 918; P < 0.05). In the AP group, the LC-DCP absorbed significantly greater energy during 10,000 cycles compared with the LCP group (P < 0.05). The 2 constructs demonstrated different biomechanical behavior with time. As cycling progressed in the LC-DCP specimens under torsion testing, stiffness (measured at the actuator at the bone ends) did not change significantly; however, fracture motion (measured at the fracture surfaces) decreased significantly (P = 0.04). The LCP specimens did not display similar behavior.

Conclusions: Our findings indicated that LCP constructs may demonstrate subtle mechanical superiority compared with the LC-DCP. The LCP specimens had less energy absorption in the AP group and survived longer in the torsion group. Discordance of motion between measurement regions was observed only in the LC-DCP torsion group, and may have been caused by plate-bone slippage or bone-screw subcatastrophic failure. However, many other compared parameters were found to be similar, and the clinical significance of the few differences found between constructs mandates further investigation.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.bot.0000174033.30054.5fDOI Listing

Publication Analysis

Top Keywords

lcp specimens
16
compression plates
12
dynamic compression
12
compression plate
12
torsion group
12
mechanical behavior
8
locking compression
8
cadaver radius
8
energy absorption
8
group
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!