We studied the tapered interface between the head and the neck of 139 modular femoral components of hip prostheses which had been removed for a variety of reasons. In 91 the same alloy had been used for the head and the stem; none of them showed evidence of corrosion. In contrast, there was definite corrosion in 25 of the 48 prostheses in which the stem was of titanium alloy and the head of cobalt-chrome. This corrosion was time-dependent: no specimens were corroded after less than nine months in the body, but all which had been in place for more than 40 months were damaged. We discuss the factors which may influence the rate of these changes and present evidence that they were due to galvanically-accelerated crevice corrosion, which was undetected in previous laboratory testing of this type of prosthesis.

Download full-text PDF

Source
http://dx.doi.org/10.1302/0301-620X.74B4.1624507DOI Listing

Publication Analysis

Top Keywords

modular femoral
8
hip prostheses
8
alloy head
8
corrosion
5
corrosion components
4
components modular
4
femoral hip
4
prostheses studied
4
studied tapered
4
tapered interface
4

Similar Publications

Objective: We present our experience with endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit (Endo-Bentall) in high-risk patients with aortic root pathologies.

Methods: The physician constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), aortic endovascular stent graft (TEVAR), and two wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft.

View Article and Find Full Text PDF

Background: Dual mobility (DM) implants in total hip arthroplasty provide excellent range of motion with low dislocation rates. A complication of this design is intraprosthetic dislocation (IPD), where the polyethylene (PE) liner dissociates from the femoral head. In older designs, IPD occurred due to a small head size and late PE wear with head-capture-mechanism failure.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to compare the Modular Dual Mobility (MDM) hip system and large femoral heads (LFHs) in revision total hip arthroplasties (THAs), focusing on risks like dislocation and metal-related complications.
  • Researchers analyzed 299 revision THAs and found that the MDM group had a significantly lower rate of re-revision for dislocation (99% survival) compared to the LFH group (91% survival), with increased risks for LFH patients.
  • Overall, THAs utilizing the MDM construct demonstrated a safer profile, particularly in reducing re-revisions related to dislocation and had no reported issues with metallosis or corrosion.
View Article and Find Full Text PDF

Objective To determine the outcomes of cemented modular bipolar hemiarthroplasty for displaced femoral neck fractures in the elderly. Methodology This prospective study involved 102 elderly patients with clinically and radiologically confirmed displaced femoral neck fractures and was conducted in the Department of Trauma and Orthopedic Surgery, Unit-1, Services Hospital, Lahore. Cemented bipolar hemiarthroplasty was performed on all patients.

View Article and Find Full Text PDF

Background: The tapered fluted titanium stem is a viable option for complex hip reconstruction. We therefore, evaluate the results of complex hip arthroplasty for femoral bone loss (Paprosky type IIIA to type IV femoral defect), using a modular fluted titanium stem.

Methods: Data for this retrospective study was collected from the medical records of the 2 participating orthopaedic units from December 2018 to December 2021.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!