Unlabelled: The objective was to design and validate a method for tele-operating (from an expert site) an echographic examination in an isolated site.

Method: The isolated places, defined as areas with reduced medical facilities, could be secondary hospitals 20 to 50 km from the university hospital, or dispensaries in Africa or Amazonia, or a moving structure like a rescue vehicle or the International Space Station (ISS). At the expert center, the ultrasound medical expert moves a fictive probe, connected to a computer (n degrees 1) which sends, the coordinate changes of this probe via an ISDN or satellite line to a second computer (n degrees 2), located at the isolated site, which applies them to the robotic arm holding the real echographic probe.

Results: The system was tested at Tours Hospital on 105 patients. A complete investigation (visualization) of all the organs requested for different clinical cases was obtained in 76% of the cases with the robot, and 87% at the reference echography: In 11% of the cases, at least one of the organ visualized at reference echo could not be investigated by the robot, thus the diagnostic was not done. The number of repositioning was higher for the robot (6.5 +/- 2) than for the reference echo (5.1 +/- 2 = or > 24% more with robot). The duration of the examination was higher with the robot (16 +/- 10 min) than for the reference echography (11 +/- 4 min = or > +43% with the robot compare to reference echography. The system was also tested successfully using satellite links in a limited number of cases (approx 30).

Download full-text PDF

Source

Publication Analysis

Top Keywords

reference echography
12
echographic examination
8
computer degrees
8
system tested
8
reference echo
8
higher robot
8
robot +/-
8
+/- min
8
robot
7
reference
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!