Purpose: To reduce radiation dose from abdominal computed tomography (CT) without degradation of low-contrast detectability by using a technique with low tube voltage (90 kV).

Materials And Methods: The institutional review board approved the participation of the radiologists in the observer performance test, and informed consent was obtained from all participating radiologists. A phantom for measurement of the radiation dose and a phantom containing low-contrast objects were scanned with a 16-detector row CT scanner at 120 kV and 90 kV. For determination of the radiation dose at both 90 kV and 120 kV, the tube current-time product settings were 100-560 mAs, and the doses at the center and periphery of the phantom were measured. To assess low-contrast detectability, we used a 300-mAs setting at 120 kV and 250-560-mAs settings at 90 kV. Five observers participated in the receiver operating characteristic analysis. Area under the receiver operating characteristic curve (A(z)) values were calculated in each observer. A(z) values obtained with each of the scanning techniques were recorded, and differences were examined for significance by using the Dunnet method.

Results: The mean A(z) value was 0.951 at 120 kV and 300 mAs. A(z) values were 0.927-0.973 at 90 kV and 450-560 mAs, and the differences between those values and values obtained at 120 kV and 300 mAs were not significant (P = .937-.952). A value of 100% was assigned to the radiation dose delivered to the center of the phantom at 120 kV and 300 mAs. The relative dose delivered at 90 kV ranged from 65% at 450 mAs to 79% at 560 mAs.

Conclusion: A reduction from 120 kV to 90 kV led to as much as a 35% reduction in the radiation dose, without sacrifice of low-contrast detectability, at CT.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2373041643DOI Listing

Publication Analysis

Top Keywords

radiation dose
24
low-contrast detectability
16
120 300
12
300 mas
12
degradation low-contrast
8
receiver operating
8
operating characteristic
8
dose delivered
8
0
7
radiation
6

Similar Publications

Purpose: Long axial field-of-view (LAFOV) positron emission tomography/computed tomography (PET/CT) scanners enable high sensitivity and wide anatomical coverage. Therefore, they seem ideal to perform post-selective internal radiation therapy (SIRT) Y scans, which are needed, to confirm that the dose is delivered to the tumors and that healthy organs are spared. However, it is unclear to what extent the use of LAFOV PET is feasible and which dosimetry approaches results in accurate measurements.

View Article and Find Full Text PDF

Dosimetric comparison of VMAT plans in preoperative short-course rectal radiotherapy.

Sci Rep

December 2024

Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, 310000, China.

Recently, neoadjuvant short-course radiation therapy (SCRT) has emerged as a valid treatment option for patients with locally advanced rectal cancer (LARC). We assessed SCRT plans using volumetric-modulated arc therapy (VMAT) with Halcyon and Infinity medical linear accelerators (Linacs) and compared the plan quality and delivery efficiency across all cases. Thirty patients who underwent preoperative SCRT for LARC at the hospital were randomly selected.

View Article and Find Full Text PDF

Radon (Rn) and thoron (Rn) were reported as the highest contributors to natural radiation received by humans. Furthermore, radon has been stated as the second-highest cause of lung cancer. The concentrations of U and Th (the parent nuclide of radon and thoron, respectively) in nature vary with geological conditions and can be enhanced by human activities.

View Article and Find Full Text PDF

Hypoxic tumors are radioresistant stemming from the fact that oxygen promotes reactive oxygen species (ROS) propagation after water radiolysis and stabilizes irradiation-induced DNA damage. Therefore, an attractive strategy to radiosensitize solid tumors is to increase tumor oxygenation at the time of irradiation, ideally above a partial pressure of 10 mm-Hg at which full radiosensitization can be reached. Historically, the many attempts to increase vascular O delivery have had limited efficacy, but mathematical models predicted that inhibiting cancer cell respiration would be more effective.

View Article and Find Full Text PDF

Optimizing the cumulative cisplatin dose for concurrent chemoradiotherapy beneficiaries among elderly nasopharyngeal carcinoma patients: a real world study.

Sci Rep

December 2024

Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital and Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen, 518116, Guangdong, China.

This study aimed to find a safe and effective cumulative cisplatin dose (CCD) for concurrent chemoradiotherapy (CCRT) beneficiaries among elderly nasopharyngeal carcinoma (NPC) patients. A total of 765 elderly (≥ 60 years old) NPC patients treated with cisplatin-based CCRT and IMRT-alone from 2007 to 2018 were included in this study. RPA-generated risk stratification was used to identify CCRT beneficiaries.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!