Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: We report a comprehensive assessment and validation of a new intraoperative angiography technique.
Background: Technical problems at the site of the distal anastomosis compromise an underappreciated proportion of coronary bypass grafts. The absence of a systematic, validated technique to verify graft patency in the operating room represents a significant breach in quality assurance.
Methods: Fluorescent indocyanine green (ICG) dye is excited with dispersed laser light to create an angiographic depiction of the graft, native vessel, and anastomosis. One-hundred twenty patients underwent ICG angiography. Angiograms were reviewed for reliability and validity studies.
Results: A total of 348 coronary bypass grafts were studied. Each ICG angiogram took 2.2 +/- 1.1 min to perform. The ICG angiography found 4.2% of patients had significant graft problems requiring major revision. Quality of visualization was rated according to a seven-point Likert scale (1 = worst, 7 = best). Among conduits, saphenous veins were best visualized (mean score +/- standard deviation), 6.4 +/- 1.5 versus 5.5 +/- 1.9 for internal mammary arteries and 4.4 +/- 2.3 for radial arteries (p = 0.02). Location of distal anastomosis did not influence quality of visualization. There was high inter-rater reliability for graft revision (kappa = 1.0) and graft patency (kappa = 0.97) between surgeons. Sensitivity and specificity of the ICG angiograms for graft stenosis >50% was 100% among 22 grafts also studied with X-ray angiography.
Conclusions: Information from ICG angiograms led to graft revisions for technical problems in 4.2% of patients that would have otherwise gone unrecognized. Intraoperative angiography is an emerging tool for improving the quality of coronary bypass surgery.