Objective: Microembolic signals detected by transcranial Doppler ultrasonography have been demonstrated to be clinically relevant in patients supported with pulsatile left ventricular assist devices. We prospectively investigated the quantity of microembolic signals in patients supported with the continuous-flow DeBakey left ventricular assist device (MicroMed DeBakey VAD; MicroMed Technology, Inc, Houston, Tex) including the refined Carmeda BioActive Surface system (Carmeda AB, Stockholm, Sweden).

Methods: Twenty-three patients (20 male) aged 14 to 62 years supported with DeBakey left ventricular assist devices (n = 6 with Carmeda) were enrolled in this study. Microembolic signal monitorings were performed twice weekly by insonating the middle cerebral artery for 20 minutes without and 20 minutes with oronasal application of oxygen (6 L/min). Evidence of clinically manifest thromboembolic events was based on regular questionnaires, clinical examinations, and results of diagnostic procedures.

Results: Despite a low incidence of thromboembolic complications (0.24 per 100 left ventricular assist device days), 20 patients (87%) showed circulating microemboli. Overall, microembolic signals were found in 175 of 499 transcranial Doppler ultrasonographic examinations (35.1%), with mean counts of 81.2 +/- 443 (range 0-5042 signals/h). Both microembolic signal prevalence (25% vs 34%, P = .01) and absolute signal counts (46.5 vs 104, P < .01) significantly declined with oxygen delivery. There was no significant correlation between the individual microembolic signal activity and the incidence of clinical thromboembolism or the intensity of antihemostatic treatment. Patients supported with the Carmeda device did not show reduced rates of clinical thromboembolization or cerebral microemboli.

Conclusion: In patients with DeBakey left ventricular assist devices, a high load of clinically silent microemboli can be detected within the cerebral arteries despite a low incidence of embolic complications. It needs to be investigated whether such continuous, presumably gaseous microembolization causes cognitive or neuropsychologic deficits.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2005.02.068DOI Listing

Publication Analysis

Top Keywords

left ventricular
24
ventricular assist
24
patients supported
16
debakey left
16
assist device
12
microembolic signals
12
assist devices
12
microembolic signal
12
supported debakey
8
transcranial doppler
8

Similar Publications

Background: Sequential application of radiofrequency with pulsed field (PF) ablation may increase lesion depth while preserving the advantages of PF. The study's aim was to determine lesion dimensions of sequential, colocalized radiofrequency and PF ablation.

Methods: A preclinical study using swine (n=4) performed lesions in the right/left ventricles.

View Article and Find Full Text PDF

Objective: Recent studies have underscored the metabolic and cardiovascular regulatory capacity of perirenal adipose tissue (PAT), implicating its potential involvement in the pathogenesis of left ventricular hypertrophy (LVH). This investigation aims to assess the relationship between increased PAT mass and LVH, while also examining the potential mediating role of insulin resistance in this relationship among individuals with type 2 diabetes mellitus (T2DM).

Method: 1112 individuals with T2DM were prospectively recruited for this study.

View Article and Find Full Text PDF

Background: This study investigates the feasibility and early outcomes of early myocardial reperfusion in patients with type A aortic dissection (TAAD), evaluating its effectiveness and potential benefits compared to traditional cardioplegic arrest techniques.

Methods: A retrospective analysis was conducted on 168 patients diagnosed with TAAD who underwent surgery at the General Hospital of the Northern Theater Command in China from January 2021 to July 2024. Patients were divided into two groups: early myocardial reperfusion (EMR group,  = 66) and cardioplegic arrest (CA group,  = 102).

View Article and Find Full Text PDF

The effect of thermoelectric craniocerebral cooling device on protecting brain functions in post-cardiac arrest syndrome.

Front Cardiovasc Med

January 2025

Department of Anesthesiology and Reanimation, Faculty of Medicine, Bezmialem Vakif University, Istanbul, Türkiye.

Aim: This study aimed to protect brain functions in patients who experienced in-hospital cardiac arrest through the application of local cerebral hypothermia. By utilizing a specialized thermal hypothermia device, this approach sought to mitigate ischemic brain injury associated with post-cardiac arrest syndrome, enhance survival rates, and improve neurological outcomes as measured by standardized scales.

Methods: A prospective, single-center cohort study was conducted involving patients aged ≥18 years who experienced in-hospital cardiac arrest and achieved return of spontaneous circulation (ROSC).

View Article and Find Full Text PDF

Objective: Chronic Chagas Cardiomyopathy (CCC) carries a high risk of embolic events due to structural changes in the left ventricle and frequent conduction disorders. However, there is limited data on anticoagulant prescription patterns and factors influencing the use of direct oral anticoagulants (DOACs) in these patients. This study aims to characterize CCC patients based on the anticoagulant therapy received and identify factors associated with DOACs use.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!