Background: The aim of this study was to compare VVI, VVIR and DDD modes in patients with indication to dual-chamber stimulation, depending on left ventricular function.
Methods: Two groups of patients were implanted with a DDD pacemaker: Group I with ejection fraction > 40% and Group II with ejection fraction < 40%. Patients with a history of atrial arrhythmia or retrograde conduction were excluded. At follow-up (1 month each) quality of life (QoL), patient preference and echo parameters were collected. At hospital discharge all patients were programmed in DDD for 1 month and then randomized to VVI or VVIR mode. At the end of the period in VVI or VVIR mode each patient underwent a control period in DDD and then was programmed in VVIR or VVI mode.
Results: Seventeen patients out of 23 preferred DDD mode and 6 did not perceive any subjective difference among DDD, VVI and VVIR modes (4/9 in Group I and 2/14 in Group II, p = 0.0017). QoL was significantly different between the two groups and at each follow-up showed the best values in DDD. The correlation between QoL and Tei index was 0.62 in Group I (p < 0.001) and 0.35 in Group II (p = 0.001). Neither ejection fraction nor fractional shortening showed any significant difference during the three phases of the study.
Conclusions: Most patients preferred the DDD mode. The Tei index showed a good correlation with QoL and both QoL and Tei index significantly improved with DDD mode as compared to VVI and VVIR.
Download full-text PDF |
Source |
---|
G Ital Cardiol (Rome)
October 2023
U.O.S.D. Aritmologia, A.O.R.N. dei Colli, Ospedale "V. Monaldi", Napoli.
Background: The Pacemaker (PM) and Implantable Cardioverter-Defibrillator (ICD) Registry of the Italian Association of Arrhythmology and Cardiac Pacing (AIAC) collects information about demographics, clinical characteristics, main indications for PM/ICD therapy and device types from the voluntary Italian collaborating centers.
Methods: For the year 2022 main data about national PM and ICD implantation activity were obtained on the basis of European Cards and subsequently analyzed to create a standard report.
Results: PM Registry: data about 17 800 PM implantations were collected (14 191 first implants and 3609 replacements).
J Cardiovasc Electrophysiol
August 2023
Cardiology Department, Clinique du Millénaire, Montpellier, France.
Introduction: The efficacy and safety of leadless cardiac pacemakers (LPMs) as an alternative to conventional transvenous cardiac pacing have been largely reported. The first generation of the Micra transcatheter pacing system (VR; Medtronic) was able to provide single-chamber VVI(R) pacing mode only, with a potential risk of pacemaker syndrome in sinus rhythm patients. A second-generation system (AV) now provides atrioventricular synchrony through atrial mechanical (Am) sensing capability (VDD mode).
View Article and Find Full Text PDFInt J Cardiol
June 2023
Tianjin Key Laboratory of Ionic-Molecular Function of Cardiovascular Disease, Department of Cardiology, Tianjin Institute of Cardiology, Second Hospital of Tianjin Medical University, Tianjin 300211, China. Electronic address:
Aims: Sick sinus syndrome (SSS) and atrial fibrillation (AF) frequently coexist and show a bidirectional relationship. This systematic review and meta-analysis aimed to decipher the precise relationship between SSS and AF, further exploring and comparing different therapy strategies on the occurrence or progression of AF in patients with SSS.
Methods And Results: A systematic literature search was conducted until November 2022.
G Ital Cardiol (Rome)
March 2023
Clinica di Cardiologia, Università dell'Insubria, Ospedale di Circolo, Varese.
Background: The Pacemaker (PM) and Implantable Cardioverter-Defibrillator (ICD) Registry of the Italian Association of Arrhythmology and Cardiac Pacing (AIAC) collects information about demographics, clinical characteristics, main indications for PM/ICD therapy and device types from the voluntary Italian collaborating centers.
Methods: For the year 2021 main data about national PM and ICD implantation activity were obtained on the basis of European Cards and subsequently analyzed to create a standard report.
Results: PM Registry: data about 18 631 PM implantations were collected (15879 first implants and 2752 replacements).
Pacing Clin Electrophysiol
April 2023
Department of Cardiology, Coimbra University Hospital Centre, Coimbra, Portugal.
Background: In elderly patients, single chamber pacing may be considered. For sinus rhythm patients, VDD pacemaker (PM), by preserving atrial sensing, is a more physiological mode than VVI devices. This study aims to evaluate the long-term performance of VDD PM in elderly patients with atrioventricular block (AVB).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!