Moria M2 single use microkeratome head in 100 consecutive LASIK procedures.

J Refract Surg

Department of Ophthalmology, Manhattan Eye, Ear and Throat Hospital, New York, NY, USA.

Published: November 2005

Purpose: To evaluate the safety and efficacy of the Moria M2 single use 130 microkeratome head in consecutive LASIK procedures for correction of myopia and myopic astigmatism.

Methods: One hundred eyes of 55 patients underwent LASIK in which the flaps were created with the Moria M2 microkeratome using the single use 130 head and excimer laser ablation was done with the Allegretto Wave-light laser. Flap parameters measured were: thickness, diameter, hinge length, and overall quality. Preoperative uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), refraction, wavefront aberrations, and low contrast sensitivity were compared to postoperative values at 6-month follow-up.

Results: Mean flap thickness was 145 +/- 17.5 microm, mean flap diameter was 8.5 +/- 0.40 mm, and mean hinge cord length was 4.05 +/- 0.35 mm. At 6-month follow-up, UCVA improved from 20/200 (+/-0.24) to 20/18.5 (+/-0.12) and BSCVA improved from 20/20.5 (+/-0.18) to 20/17.5 (+/-0.11).

Conclusions: The Moria M2 single use 130 microkeratome head appears to be safe and effective in performing LASIK procedures.

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081-597X-20050901-09DOI Listing

Publication Analysis

Top Keywords

moria single
12
microkeratome head
12
lasik procedures
12
single 130
12
consecutive lasik
8
130 microkeratome
8
visual acuity
8
moria
4
microkeratome
4
single microkeratome
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!