Background: Little is known about the accuracy of capsule endoscopy (CE) in evaluation of small-bowel Crohn's disease.

Methods: Symptomatic eligible patients had ileocolonoscopy and biopsies from the terminal ileum, followed by small-bowel radiologic studies before CE. Endoscopic, radiologic, CE, and histologic findings were compared. Histology (terminal ileum biopsy specimens or a tissue sample after small-bowel resection) served as a criterion standard.

Results: Fifty-four patients were enrolled; 15 of the 54 patients were excluded from data analysis (critical small-bowel strictures, 14, identified on radiology; incomplete CE, 1). Data were analyzed for 39 patients. All patients had histologic evaluation of the small bowel. Final diagnosis of active small-intestine Crohn's disease was made in 29/39 patients (74.4%). When calculated, CE yielded a sensitivity and a specificity of 89.6% and 100.0%, respectively, and a positive predictive value and a negative predictive value of 100.0% and 76.9%, respectively, whereas small-bowel series were 27.6%/100.0% and 100.0%/32.3%.

Conclusions: CE is more accurate in detecting small-bowel inflammatory changes suggestive of Crohn's disease than conventional studies. CE, combined with ileocolonoscopy, may be proposed as a first-line investigation of the small intestine in cases of uncomplicated known or suspected Crohn's disease.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.gie.2005.06.026DOI Listing

Publication Analysis

Top Keywords

crohn's disease
16
capsule endoscopy
8
small-bowel crohn's
8
histologic findings
8
terminal ileum
8
patients
7
small-bowel
7
crohn's
5
endoscopy findings
4
findings patients
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!