Clinical performance of a ventricular automatic capture verification algorithm.

Pacing Clin Electrophysiol

Kerckhoff Klinik, Benekestrasse 2-8, D-61231 Bad Nauheim, Germany.

Published: September 2005

Background: This study was conducted to evaluate the clinical performance of the ventricular automatic capture feature as implemented in the Insignia I Ultra pacemaker system (Guidant) utilizing a variety of ventricular leads. Currently, the optimal programming of the pacemaker output considers both pacemaker efficiency (prolonging battery longevity) and patient safety (adequate safety margin). The ability of a pacemaker to automatically adjust the ventricular output above the pacing threshold while maintaining the appropriate safety margin has been explored since the early 1970s and is only available today in conjunction with a specific low polarization lead system.

Methods: One hundred and five patients were enrolled from 17 European centers utilizing 31 different types of ventricular leads were followed through their 3-month follow-up visit. There were no restrictions on the type of ventricular leads used.

Results: The average mean difference between the commanded autothreshold test (0.652 + 0.335 V) and the manual threshold test (0.651 + 0.335 V) was 0.001 + 0.49 (P < 0.0001). The average mean difference between the ambulatory autothreshold test (0.696 + 0.322 V) and the commanded autothreshold test (0.682 + 0.315 V) was 0.002 + 0.74 (P < 0.0001). Holter recordings confirmed that there were no loss of capture incidences without a backup pulse being delivered. In addition, the mean number of backup pulses delivered in a 24-hour period was less than 0.1% of the total number of paced beats.

Conclusions: This study provided that the automatic capture feature while using a variety of leads accurately determines the ventricular stimulation threshold and safely delivers a backup pulse when required.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2005.00196.xDOI Listing

Publication Analysis

Top Keywords

automatic capture
12
ventricular leads
12
autothreshold test
12
clinical performance
8
performance ventricular
8
ventricular automatic
8
capture feature
8
safety margin
8
average difference
8
commanded autothreshold
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!