Screening method for benzodiazepines and hypnotics in hair at pg/mg level by liquid chromatography-mass spectrometry/mass spectrometry.

J Chromatogr B Analyt Technol Biomed Life Sci

Laboratoire ChemTox, rue Grüninger, 67400 Illkirch, France.

Published: October 2005

A procedure is presented for the screening of 16 benzodiazepines and hypnotics in human hair by LC-MS/MS (alprazolam, 7-aminoclonazepam, 7-aminoflunitrazepam, bromazepam, clobazam, diazepam, lorazepam, lormetazepam, midazolam, nordiazepam, oxazepam, temazepam, tetrazepam, triazolam, zaleplon and zolpidem). The method involves decontamination of hair with methylene chloride, hair cut into small pieces, incubation of 20 mg in phosphate buffer (pH 8.4) in the presence of 1 ng diazepam-d5 used as internal standard, liquid-liquid extraction with diethyl ether/methylene chloride (10/90) and separation using liquid chromatography-tandem mass spectrometry. The limits of quantification for all benzodiazepines and hypnotics range from 0.5 to 5 pg/mg using a 20-mg hair sample. Linearity is observed from the limit of quantification of each compound to 200 pg/mg (r2 > 0.99). Coefficients of variation measured on six points and at two concentrations (10 and 50 pg/mg) range from 5 to 20% for all drugs but one. Extraction recovery, measured at the two same concentrations range from 32 to 76%. These results were found suitable to screen for 16 benzodiazepines in hair and detect them at very low concentrations, making this method suitable to monitor single dose.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jchromb.2004.12.036DOI Listing

Publication Analysis

Top Keywords

benzodiazepines hypnotics
12
hair
6
screening method
4
benzodiazepines
4
method benzodiazepines
4
hypnotics hair
4
pg/mg
4
hair pg/mg
4
pg/mg level
4
level liquid
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!