A femoral artery cannula that allows distal blood flow.

J Thorac Cardiovasc Surg

Department of Cardiac Surgery Research, Allegheny General Hospital, Pittsburgh, PA 15212, USA.

Published: September 2005

Objective: A femoral artery cannula is used for certain types of circulatory support but can cause ischemia, especially during prolonged perfusion. This study tests the function of a femoral cannula designed to allow proximal and distal blood flow.

Methods: Five pigs were used in the study. In each animal a distal-flow cannula was implanted in the femoral artery of one leg, and the same-sized standard cannula was implanted in the other. Blood was drained from the left atrium and delivered to the femoral artery through the distal-flow cannula or standard cannula by using a centrifugal pump. An ultrasonic flow probe and microspheres were used to quantify flow and perfusion distal to the cannula.

Results: Distal femoral flow and tissue perfusion were present in all animals (5/5) with the distal-flow cannula but only in 1 of 5 animals with the standard cannula (P < .048). Distal flow did not change with pump flow. Mean distal flow at each level of pump flow was higher with the distal-flow cannula (P < .05). Tissue perfusion was also higher with the distal-flow cannula (0.052 +/- 0.028 vs 0.010 +/- 0.022 mL x min(-1) x g(-1), P < .03).

Conclusions: In the swine model the distal-flow cannula allowed greater and more consistent distal flow than the standard cannula. The use of a distal-flow cannula for circulatory support might reduce the risk of distal limb ischemia.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2005.03.022DOI Listing

Publication Analysis

Top Keywords

distal-flow cannula
28
femoral artery
16
standard cannula
16
cannula
14
distal flow
12
flow
9
artery cannula
8
distal
8
distal blood
8
circulatory support
8

Similar Publications

Venoarterial (VA) extracorporeal membrane oxygenation (ECMO) is a treatment strategy for pediatric patients with cardiopulmonary failure. Femoral cannulation is a commonly used technique for obtaining central access in children and adolescents despite high rates of vascular complications. Ischemic limb complications remain one of the most common problems facing survivors of femoral VA ECMO.

View Article and Find Full Text PDF

Objective: Femoral arterial cannulation is associated with a significant risk of lower limb ischemia. The aim of the study was to assess the pressure and flow in the femoral artery using a novel bidirectional femoral cannula in a sheep model of peripheral cardiopulmonary bypass.

Methods: Peripheral cardiopulmonary bypass was established using a multistage venous cannula inserted into the internal jugular vein and the bidirectional or a conventional arterial cannula into the femoral artery in seven adult ewes.

View Article and Find Full Text PDF

Background: Isovolumic acceleration (IVA) obtained by tissue Doppler echocardiography (TDE) is a sensitive and relatively load-independent index for assessing systolic ventricular function. IVA also has the ability to describe the force-frequency relationship during incremental atrial pacing in vivo.

Objective: We sought to assess the ability of IVA to detect global left ventricular (LV) dysfunction induced by coronary constriction.

View Article and Find Full Text PDF

Myocardial salvage through coronary sinus intervention has been documented. The AutoRetroPerfusion Cannula is a novel device that is able to perfuse the coronary bed retrogradely through the coronary sinus with arterial blood generated from a peripheral artery with no need for a pump. The cannula consists of a distal end that, once secured in the coronary sinus, opens an umbrella-like membrane to create pressure in the coronary sinus, and at the same time has small channels directed backwards to the right atrium to provide pressure relief.

View Article and Find Full Text PDF

A femoral artery cannula that allows distal blood flow.

J Thorac Cardiovasc Surg

September 2005

Department of Cardiac Surgery Research, Allegheny General Hospital, Pittsburgh, PA 15212, USA.

Objective: A femoral artery cannula is used for certain types of circulatory support but can cause ischemia, especially during prolonged perfusion. This study tests the function of a femoral cannula designed to allow proximal and distal blood flow.

Methods: Five pigs were used in the study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!