A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Comparison of recirculation percentage of the palindrome catheter and standard hemodialysis catheters in a swine model. | LitMetric

Comparison of recirculation percentage of the palindrome catheter and standard hemodialysis catheters in a swine model.

J Vasc Interv Radiol

Yale Center for Vascular Disease, Yale University School of Medicine, SP Room 323, 20 York Street, New Haven, Connecticut 06504, USA.

Published: September 2005

Purpose: The purpose of this study was to evaluate in vivo recirculation of a new bi-directional dialysis catheter with a symmetric tip design and compare it to currently marketed staggered and split tip catheters.

Materials And Methods: Twelve catheters were placed in three 70-kg male swine. The catheters were connected to a roller type peristaltic pump. Blood flow and recirculation rates were measured with ultrasound (US) dilution. Recirculation was measured in blood flow rates of 300 mL/min, 400 mL/min, and 500 mL/min. The dialysis lines were then reversed and recirculation was measured in flow rate of 400 mL/min. Measurements were conducted with the catheter tip in the right atrium (RA) and repeated with the catheter tip in the superior vena cava (SVC).

Results: Recirculation rates measured with the Palindrome catheter tip (Tyco Healthcare Group LP, Mansfield, MA) in the SVC and in the RA were 0%. When the dialysis lines were reversed, the Palindrome catheter average recirculation was 0% with the tip in the SVC and 1% (0%-3%) with the tip in the RA. The average recirculation rates measured with the split tip catheter were 1% (0%-10%) in the SVC and 0% when the catheter tip was in the RA. When the dialysis lines were reversed split tip catheter recirculation was 11% (9%-16%) with the tip in the SVC and 13% (7%-18%) with the tip in the RA. With the staggered tip catheter, 1% (0%-5%) recirculation was measured. When the dialysis lines were reversed, the average recirculation in the staggered tip catheter measured 23% (18%-30%). Recirculation rate with reversal of dialysis lines was significantly lower with the Palindrome catheter when compared to the split tip and staggered tip designs (P < .05).

Conclusions: Preliminary results in vivo indicate that the Palindrome catheter demonstrates minimal recirculation with reversal of the dialysis lines compared to currently marketed split tip and staggered tip catheters.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.RVI.0000171700.45582.9EDOI Listing

Publication Analysis

Top Keywords

dialysis lines
24
palindrome catheter
20
lines reversed
16
catheter
13
recirculation
12
recirculation rates
12
rates measured
12
recirculation measured
12
average recirculation
12
currently marketed
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!