Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: The purpose of this study was to evaluate in vivo recirculation of a new bi-directional dialysis catheter with a symmetric tip design and compare it to currently marketed staggered and split tip catheters.
Materials And Methods: Twelve catheters were placed in three 70-kg male swine. The catheters were connected to a roller type peristaltic pump. Blood flow and recirculation rates were measured with ultrasound (US) dilution. Recirculation was measured in blood flow rates of 300 mL/min, 400 mL/min, and 500 mL/min. The dialysis lines were then reversed and recirculation was measured in flow rate of 400 mL/min. Measurements were conducted with the catheter tip in the right atrium (RA) and repeated with the catheter tip in the superior vena cava (SVC).
Results: Recirculation rates measured with the Palindrome catheter tip (Tyco Healthcare Group LP, Mansfield, MA) in the SVC and in the RA were 0%. When the dialysis lines were reversed, the Palindrome catheter average recirculation was 0% with the tip in the SVC and 1% (0%-3%) with the tip in the RA. The average recirculation rates measured with the split tip catheter were 1% (0%-10%) in the SVC and 0% when the catheter tip was in the RA. When the dialysis lines were reversed split tip catheter recirculation was 11% (9%-16%) with the tip in the SVC and 13% (7%-18%) with the tip in the RA. With the staggered tip catheter, 1% (0%-5%) recirculation was measured. When the dialysis lines were reversed, the average recirculation in the staggered tip catheter measured 23% (18%-30%). Recirculation rate with reversal of dialysis lines was significantly lower with the Palindrome catheter when compared to the split tip and staggered tip designs (P < .05).
Conclusions: Preliminary results in vivo indicate that the Palindrome catheter demonstrates minimal recirculation with reversal of the dialysis lines compared to currently marketed split tip and staggered tip catheters.