  • Dexsil 300 and Dexsil 400 are specialized copolymers made from bis(dimethylsilyl)-m-carborane and different siloxanes, designed for use in gas chromatography.
  • Their gas chromatographic performance was assessed using specific metrics, revealing the bis(dimethylsilyl)-m-carborane component's strong influence on selectivity and temperature behavior.
  • The polymers demonstrated good stability with low column bleeding levels even at high temperatures, with Dexsil 300 and 400 operating effectively up to 380 degrees Celsius.

Article Abstract

The carborane-siloxane copolymers Dexsil 300, a 34.5% bis(dimethylsilyl)-m-carborane-65.5% dimethylsiloxane copolymer, and Dexsil 400, a 24.9% bis(dimethylsilyl)-m-carborane-50.8% dimethyl, 24.3% methylphenylsiloxane copolymer, were coated on fused silica capillary columns and their gas chromatographic properties were evaluated. Their selectivity was evaluated using both Rohrschneider-McReynolds constants and triacylglycerol indices. The bis(dimethylsilyl)-m-carborane unit turned out to be equivalent to two dimethylsiloxy units and one half of a diphenylsiloxy unit. The m-carborane unit was found to cause a 15-25 K shift in the elution temperature between 120 and 360 degrees C. The working range was from 20 and 0 degrees C to 380 degrees C for Dexsil 300 and Dexsil 400, respectively. The column bleeding levels at 380 degrees C were below 20 and 15 pA for Dexsil 300 and Dexsil 400, respectively.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00216-005-3399-6DOI Listing

Publication Analysis

Top Keywords

dexsil 300
12
dexsil 400
12
capillary columns
8
380 degrees
8
degrees dexsil
8
300 dexsil
8
dexsil
6
preparation characterization
4
characterization fused-silica
4
fused-silica capillary
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!