A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Variability among five over-the-counter blood glucose monitors. | LitMetric

Variability among five over-the-counter blood glucose monitors.

Clin Chim Acta

Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Mailstop F25, 4770 Buford Hwy NE, Atlanta, GA 30341-3724, USA.

Published: February 2006

Background: The American Diabetes Association recommends that people with diabetes use self-monitoring to control their blood glucose concentration. To assess the need for standardization, we evaluated the variability among 5 of the most common monitors: MediSense Precision Xtra, Ascencia Dex, Prestige Smart System, OneTouch Ultra, and Accu-Chek Advantage.

Methods: We took steps to minimize preanalytical variation. We also eliminated user variability by using one trained operator to collect samples and perform all testing. Each monitor was used twice with each participant; one test was performed using an aged strip and the other using a fresh strip. We compared monitors using a separate ANOVA for each concentration range and strip lot.

Results: The total CVs and the within-strip lot CVs were not statistically different among monitors, ranging from 3.1% to 11.3% and from 2.1% to 8.5%, respectively. There were statistically significant differences among monitors for among-strip lot CVs, which ranged from nearly 0% to 7.5%. The degree of significance increased as the concentration range increased [3.9-5.5 mmol/l: p<0.05; 5.6-7.7 mmol/l: p =0.003; 7.8-11.1 mmol/l: p < 0.001]. The average percent difference between monitor pairs was statistically significant (p < 0.05) in more than half of the paired comparisons, with significant differences ranging from 5.7% to 32.0%.

Conclusions: Monitor results can vary significantly so that agreement among them is poor. Standardization is necessary to minimize variability and to improve patient care.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2005.07.027DOI Listing

Publication Analysis

Top Keywords

blood glucose
8
concentration range
8
lot cvs
8
monitors
5
variability over-the-counter
4
over-the-counter blood
4
glucose monitors
4
monitors background
4
background american
4
american diabetes
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!