Purpose: To prospectively compare dose reduction and image quality achieved with an automatic exposure control system that is based on both angular (x-y axis) and z-axis tube current modulation with dose reduction and image quality achieved with an angular modulation system for multi-detector row computed tomography (CT).

Materials And Methods: The study protocol was approved by the institutional review board, and oral informed consent was obtained. In two groups of 200 patients, five anatomic regions (ie, the thorax, abdomen-pelvis, abdomen-liver, lumbar spine, and cervical spine) were examined with this modulation system and a six-section multi-detector row CT scanner. Data from these patients were compared with data from 200 patients who were examined with an angular modulation system. Dose reduction by means of reduction of the mean effective tube current in 600 examinations, image noise in 200 examinations performed with each modulation system, and subjective image quality scores in 100 examinations per-formed with each modulation system were compared with Wilcoxon signed rank tests.

Results: Mean dose reduction for the angular and z-axis tube current modulation system and for the angular modulation system was as follows: thorax, 20% and 14%, respectively; abdomen-liver, 38% and 18%, respectively; abdomen-pelvis, 32% and 26%, respectively; lumbar spine, 37% and 10%, respectively; and cervical spine, 68% and 16%, respectively. These differences were statistically significant (P < .05). There was no significant difference in image noise and mean image quality scores between modulation systems, with the exception of cervical spinal examinations (P < .001 for both), where the examinations with angular modulation resulted in better scores. There is good correlation between the mean effective tube current level and the body mass index of patients with the new modulation system. Correlation was as follows: thorax, 0.77; abdomen-pelvis, 0.83; abdomen-liver, 0.84; lumbar spine, 0.8; and cervical spine, 0.6. This correlation was not observed with the angular modulation system.

Conclusion: An automatic exposure control mechanism that is based on real-time anatomy-dependent tube current modulation delivers good image quality with a significantly reduced radiation dose.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2363041220DOI Listing

Publication Analysis

Top Keywords

modulation system
32
image quality
20
tube current
20
angular modulation
20
dose reduction
16
modulation
13
automatic exposure
12
exposure control
12
multi-detector row
12
current modulation
12

Similar Publications

Background: Psychosis, marked by detachment from reality, includes symptoms like hallucinations and delusions. Traditional herbal remedies like kratom are gaining attention for psychiatric conditions. This was aimed at comprehending the molecular mechanisms of Kratom's antipsychotic effects utilizing a multi-modal computational approach.

View Article and Find Full Text PDF

Probiotics Exert Gut Immunomodulatory Effects by Regulating the Expression of Host miRNAs.

Probiotics Antimicrob Proteins

January 2025

Department of Reproductive Medicine, The Affiliated Hospital, Southwest Medical University, Luzhou, 646000, China.

Probiotics exert a diverse range of immunomodulatory effects on the human gut immune system. These mechanisms encompass strengthening the intestinal mucosal barrier, inhibiting pathogen adhesion and colonization, stimulating immune modulation, and fostering the production of beneficial substances. As a result, probiotics hold significant potential in the prevention and treatment of various conditions, including inflammatory bowel disease and colorectal cancer.

View Article and Find Full Text PDF

Phased structures for lossless ion manipulation offer significant improvements over the scanning second gate method for coupling with ion trap mass analyzers. With an experimental run time of under 1 min for select conditions and an average run time of less than 4 min, this approach significantly reduces experimental time while enhancing the temporal duty cycle. The outlined SLIM system connects to an ion trap mass analyzer via a PCB stacked ring ion guide, which replaces the commercial ion optics and capillary inlet.

View Article and Find Full Text PDF

Parkinson's Disease (PD) is a neurodegenerative disorder marked by the depletion of dopaminergic neurons. Recent studies highlight the gut-liver-brain (GLB) axis and its role in PD pathogenesis. The GLB axis forms a dynamic network facilitating bidirectional communication between the gastrointestinal tract, liver, and central nervous system.

View Article and Find Full Text PDF

Hematopoietic stem cell transplantation (HSCT) is a highly effective therapy for malignant blood illnesses that pose a high risk, as well as diseases that are at risk due to other variables, such as genetics. However, the prevalence of graft-versus-host disease (GVHD) has impeded its widespread use. Ensuring the stability of microbial varieties and associated metabolites is crucial for supporting metabolic processes, preventing pathogen intrusion, and modulating the immune system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!