Background: Following an accidental observation of reduced sensitivity for detection of antibodies to hepatitis C virus (HCV) with a novel commercially available automated chemiluminescent microparticle immunoassay (CMIA, ARCHITECT Anti-HCV) compared to a well-established microparticle enzyme immunoassay (MEIA, AxSYM HCV Version 3.0), we wanted to explore whether this could be explained by a variation in marginal sensitivity, to be expected between highly sensitive assays of different formats, or represented a reduced sensitivity of the CMIA to certain antibody profiles.
Objectives: To evaluate the ability of the CMIA to detect low concentrations of anti-HCV antibodies as defined by various patterns in the recombinant immunoblot assay (RIBA) and already detected in the MEIA system.
Study Design: All patient sera tested for anti-HCV reactivity during a period of 3 years (27,978) were evaluated. A total of 90 sera had a sample/cut-off ratio (S/CO) between 1.0 and 1.5 in the MEIA test and were available for further testing. Of these, 19 had a probable/possible presence of anti-HCV antibodies based on presence of at least two bands of > or = 1+ strength in the RIBA, or because the patient was known to be anti-HCV positive. These 19 sera were tested with the CMIA. In addition, 16 sera with strong reactivity to various antigen combinations in the RIBA were serially diluted until testing negative in both microparticle test systems.
Results: Seven of the 19 sera (37%) were negative (S/CO < 1.0) in the CMIA. At least 3 (16%) of these 19 sera were very likely to be true anti-HCV positive sera (from infants with known anti-HCV positive mothers). HCV-RNA was not detected in any of the sera tested. Testing of sera after serial dilution indicates that the CMIA has a lower sensitivity to c22- and c33c-antibodies than the MEIA, possibly also to c100-3-antibodies.
Conclusion: Our findings indicate that ARCHITECT Anti-HCV is less sensitive than AxSYM HCV Version 3.0 in detecting antibodies to c22 and c33c in patients who have cleared their HCV-infection and have naturally declining levels of antibodies.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcv.2005.05.013 | DOI Listing |
J Clin Virol
December 2024
Abbott Diagnostics, Abbott Park, IL, USA. Electronic address:
Introduction: HCV antibody assays have been used to screen for HCV, but confirmation of acute infection is dependent on RNA or core antigen testing. The aim of the study was to compare the performance of five HCV test methods, including RNA testing, on a US emergency department population.
Methods: Clinical performance metrics were calculated on 708 consenting Johns Hopkins Emergency Department patients who self-reported an increased risk for HCV infection.
Diagn Microbiol Infect Dis
October 2024
Director of Client Services & Technology, Westgard QC, Inc., 7614 Gray Fox Trail, Madison, WI, 53717, USA.
Sigma metric analysis was conducted across two New Zealand Blood Services (NZBS) laboratories (Auckland and Christchurch) to optimize quality control (QC) procedures. We evaluated five assays (anti-HCV, HIV Ag/Ab combo, HTLV-I/II, HBsAg, and Syphilis) using internal quality control (IQC) and third-party daily QC data extracted from four Architect i2000SR instruments during Jan 2 -31st, 2023. Mean, standard deviation (SD), and coefficient of variation (CV%) were calculated, assuming zero bias.
View Article and Find Full Text PDFThe Westgard quality control (QC) rules are often applied in infectious diseases serology to validate the quality of results, but this requires a reasonable tradeoff between maximum sensitivity to errors and minimum false rejections. This article, in addition to illustrate the six sigma methodology in the QC management of the (anti-HCV Architect®) test, it discusses the main influencing factors on sigma value. Data from low positive and in-kit control materials spreading over 6 months and using four reagent kits, were used to calculate the precision of the test.
View Article and Find Full Text PDFViruses
March 2024
Laboratory of Virology, National Institute for Infectious Diseases Lazzaro Spallanzani IRCCS, 00149 Rome, Italy.
We investigated the frequency and serological correlates of occult hepatitis B virus infection (OBI) and the potential impact of a highly sensitive assay for HBsAg in subjects infected by human immunodeficiency virus (HIV) or hepatitis C virus (HCV), who are also at risk for hepatitis B virus (HBV) infection, often in an occult form. Samples from 499 patients with HIV, all HBsAg negative and anti-HBc positive, and 137 patients with HCV were tested for HBV-DNA, anti-HBc, anti-HBs, and HBsAg by a conventional and highly sensitive assay. HBV biomarkers were detected in 71.
View Article and Find Full Text PDFJ Infect Dev Ctries
January 2024
Erzincan Binali Yıldırım University, Faculty of Medicine, Department of Medical Microbiology, Erzincan, Turkey.
Introduction: The measurement of hepatitis C virus (HCV) RNA is a test that requires high cost, advanced technique, and qualified personnel. Diagnosis and treatment of patients may be delayed due to the high rate of false-positive results. This study aims to predict true antibody positivity and viremia by determining the most appropriate anti-HCV signal-to-cutoff (S/Co) value reflecting HCV infection.