This paper presents a microfluidic chip capable of performing precise continuous pH measurements in an automatic mode. The chip is fabricated using micro-electro-mechanical-systems (MEMS)-based techniques and incorporates polydimethylsiloxane (PDMS) microstructures, pH-sensing electrodes and pneumatic fluid-control devices. Through its enhanced microchannel design and use of pneumatic fluid-control devices, the microfluidic chip reduces the dead volume of the sample and increases the pumping rate. The maximum pumping rate of the developed micro-pump is 28 microL/min at an air pressure of 10 psi and a driving frequency of 10 Hz. The total sample volume consumed in each sensing operation is just 0.515 microL. As a result, the developed chip reduces the sample volume compared to conventional large-scale pH-sensing systems. The microfluidic chip employs the electrochemical sensing method to conduct precise pH level measurements. The sensing electrodes are fabricated by sputtering a layer of SiO(2)-LiO(2)-BaO-TiO(2)-La(2)O(3) (SLBTLO) onto platinum (Pt) electrodes and the pH value of the sample is evaluated by measuring the potential difference between the sensing electrodes and a reference electrode. Additionally, the integration of the microfluidic chip with a pneumatic fluid-control device facilitates automatic sample injection and a continuous sensing operation. The developed system provides a valuable tool with which to examine pH values in a wide range of biomedical and industrial applications.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.bios.2005.06.005DOI Listing

Publication Analysis

Top Keywords

pneumatic fluid-control
16
microfluidic chip
16
fluid-control devices
12
chip reduces
8
pumping rate
8
sample volume
8
sensing operation
8
sensing electrodes
8
chip
6
microfluidic
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!