106 unicondylar knee replacement tibial components were retrieved and analyzed for the amount and type of polyethylene wear. Three different designs were retrieved which had essentially the same femorotibial conformity. Each design showed a characteristic failure pattern. The polyethylene of PCA tibial components showed serious delamination after only short durations, as a result of heat pressing. St Georg sledge prostheses showed some delamination after 4 years' duration due to sub-surface cracks which were initiated by fusion defects in the polyethylene; metal backing of the components did not affect delamination of this prosthesis. The Marmor designs showed the least wear, with shiny depressions and surface pitting; no delamination was observed in the Marmor prosthesis. Molecular weight determination by gel permeation chromatography and analysis of crystallinity using Fourier transformation infra-red spectroscopy demonstrated that St George polyethylene had higher molecular weight and crystallinity than Marmor polyethylene. In some of the components investigated, crystallinity and molecular weight of the polyethylene were reduced under the wear track when compared with the unworn polyethylene. Since fusion defects may cause delamination of polyethylene we urge manufacturers to reduce the number of such defects.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3109/17453679209154776 | DOI Listing |
J Arthroplasty
January 2025
Georgetown University School of Medicine, MedStar Georgetown University Hospital, Washington, D.C, USA. Electronic address:
Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.
View Article and Find Full Text PDFJ Clin Med
January 2025
Department of Mechanical Engineering, Yonsei University, Seoul 03722, Republic of Korea.
: Patient-specific instrumentation (PSI) in total knee arthroplasty (TKA) uses preoperative three-dimensional imaging to create cutting blocks tailored to patient anatomy. However, there is debate regarding the additional benefits of PSI in terms of improved alignment or functional outcomes compared to using conventional instruments. Although PSI design has undergone continuous development, the improvements have not been incorporated.
View Article and Find Full Text PDFClin Orthop Relat Res
January 2025
Department of Rehabilitation, Radboud University Medical Center, Nijmegen, the Netherlands.
Background: Many patients with a lower limb socket-suspended prothesis experience socket-related problems, such as pain, chronic skin conditions, and mechanical problems, and as a result, health-related quality of life (HRQoL) is often negatively affected. A bone-anchored prosthesis can overcome these problems and improve HRQoL, but these prostheses have potential downsides as well. A valid and reliable tool to assess potential candidates for surgery concerning a favorable risk-benefit ratio between potential complications related to bone-anchored prostheses and improvements in HRQoL is not available yet.
View Article and Find Full Text PDFKnee Surg Sports Traumatol Arthrosc
January 2025
Clinical Research Department, Sydney Knee Specialists, Kogarah, New South Wales, Australia.
Purpose: In functionally aligned (FA) total knee arthroplasty (TKA), femoral component rotation (FCR) is personalised to optimise flexion gap balance. As axial malalignment has been attributed to patellofemoral complications, this study assessed FA FCR in relation to the surgical transepicondylar axis (TEA) and early implant survivorship.
Methods: We analysed 446 robotic-assisted primary TKAs in 393 patients using FA with preresection gap balancing.
BMC Musculoskelet Disord
January 2025
Senior Department of Orthopedics, The Fourth Medical Center of PLA General Hospital, No. 51 Fucheng Road, Beijing, 100048, People's Republic of China.
Background: In medial mobile-bearing unicompartmental knee arthroplasty (MB-UKA), the position of the bearing does not correspond to the planned position which will increasing the risk of bearing dislocation. This study aimed to explore the relationship between the malposition of the femoral and tibial components and the phenomenon of bearing deviation using postoperative radiological measurements.
Methods: One hundred twenty patients who underwent mobile-bearing uni-compartmental knee arthroplasty (MB-UKA) at our hospital between January and August 2023 were enrolled in this retrospective study.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!