Purpose: To prospectively compare the accuracy of 16-detector row computed tomographic (CT) angiography with conventional digital subtraction angiography (DSA) as the reference standard in the assessment of aortoiliac and lower extremity arteries in patients with peripheral arterial disease.

Materials And Methods: This study was approved by the institutional review board, and informed consent was obtained. A total of 39 consecutive patients (27 men [mean age, 66 years] and 12 women [mean age, 64 years]) with peripheral arterial disease underwent both conventional DSA and 16-detector row CT angiography. For data analysis, the arterial vascular system was divided into 35 segments. A total of 1365 arterial segments were analyzed for arterial stenosis by two independent blinded readers using a four-point grading system (grade 1, <10% luminal narrowing; grade 2, 10%-49% luminal narrowing; grade 3, 50%-99% luminal narrowing; grade 4, occlusion). Interobserver agreements were calculated by using kappa statistics. A third independent blinded reader assessed possible reasons for disagreements between 16-detector row CT angiographic findings and conventional DSA findings. Effective radiation dose was calculated for both imaging modalities.

Results: Sixteen-detector row CT angiographic and conventional DSA findings were diagnostic in all vascular segments. Compared with conventional DSA, the sensitivity and specificity of 16-detector row CT angiography with regard to detection of hemodynamically significant stenosis in all 35 arterial segments were 96% and 97%, respectively, for both readers. Readers 1 and 2 overestimated arterial stenosis in 42 (3%) and 34 (2%) arterial segments, respectively, and underestimated arterial stenosis in 13 (1%) and 10 (1%) arterial segments, respectively. Interobserver agreement was excellent (kappa = 0.84-1.00). Presence of anteroposteriorly located luminal narrowing and extensive vascular wall calcification were considered main reasons for disagreements between imaging modalities. Effective radiation dose was lower for 16-detector row CT angiography (1.6-3.9 mSv) than for conventional DSA (6.4-16.0 mSv).

Conclusion: Sixteen-detector row CT angiography is an accurate and reliable noninvasive alternative to conventional DSA in the assessment of aortoiliac and lower extremity arteries in patients with peripheral arterial disease.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiol.2362040895DOI Listing

Publication Analysis

Top Keywords

16-detector row
12
aortoiliac lower
8
lower extremity
8
extremity arteries
8
row angiography
8
digital subtraction
8
subtraction angiography
8
peripheral arterial
8
[mean age
8
age years]
8

Similar Publications

Volumetric Analysis of Carotid Plaque Components and Cerebral Microbleeds: A Correlative Study.

J Stroke Cerebrovasc Dis

March 2017

Point-of-Care Devices, Global Biomedical Technologies, Inc., Roseville, CA, USA; Stroke Monitoring and Diagnostic Division, AtheroPoint™, Roseville, CA, 95661, USA; Department of Electrical Engineering, University of Idaho (Affl.), Pocatello, ID, USA.

Purpose: The purpose of this work was to explore the association between carotid plaque volume (total and the subcomponents) and cerebral microbleeds (CMBs).

Materials And Methods: Seventy-two consecutive (male 53; median age 64) patients were retrospectively analyzed. Carotid arteries were studied by using a 16-detector-row computed tomography scanner whereas brain was explored with a 1.

View Article and Find Full Text PDF

Objective: To determine the lesion characterization capability by low dose CT for localized lung lesions in comparison with standard dose CT.

Subjects And Methods: Approval for this study was granted by our Institutional Review Board. Fifty-two consecutive patients (36 males and 16 females, median age of 71 years.

View Article and Find Full Text PDF

Objectives: This study aimed to evaluate the diagnostic accuracy of coronary computed tomography angiography (CTA) for detecting coronary artery stenoses in patients with valvular heart disease undergoing valve surgery.

Background: Coronary CTA is currently not routinely recommended for detecting coronary artery stenoses before cardiac valve surgery. However, recent improvements in computed tomography technology may enable the identification of the most appropriate candidates for coronary CTA before valve surgery.

View Article and Find Full Text PDF

Diagnostic Impact of Bone-Subtraction CT Angiography for Patients with Acute Subarachnoid Hemorrhage.

AJNR Am J Neuroradiol

February 2016

From the Department of Neuroradiology (P.A., K.E., K.H., R.v.K.), University Hospital Carl Gustav Carus, Technische Universität, Dresden, Germany

Background And Purpose: Detection and evaluation of ruptured aneurysms is critical for choosing an appropriate endovascular or neurosurgical intervention. Our aim was to assess whether bone-subtraction CTA is capable of guiding treatment for cerebral aneurysms in patients with acute SAH and could replace DSA.

Materials And Methods: We prospectively studied 116 consecutive patients with SAH with 16-detector row bone-subtraction CTA and DSA before intracranial aneurysm treatment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!