Background: State-of-the art cardiac pacemakers are protected against radiofrequency signals. Although there have been earlier clinical and in vitro reports of cellular phone interference with implantable devices, only a few studies have been performed in recent years. The ringing phase of digital GSM or PCS cellular phones includes a brief period of peak radiated power.

Objectives: This study tested the protection offered by electromagnetic filters of cardiac pacemakers against cellular phone ringing.

Methods: We performed 330 consecutive tests in 158 patients at the time of routine examination in our pacemaker follow-up clinic. The programmed parameters remained unchanged before testing. During electrocardiographic monitoring, 2 single-band digital cellular phones consecutively placed over the pacemaker pocket each received a call. The phone systems tested were 1) GSM at a maximal power output of 2 W, operating on a 900 MHz carrier frequency, and 2) PCS at a maximal output of 1 W, operating on a 1800 MHz carrier frequency.

Results: Interference was noted in only 5 tests, due to interaction by the GSM system with 4 unprotected pacemaker models. The GSM test was negative in 12 other tests of identical pulse generator models. The overall incidence of interference was 1.5% of tests.

Conclusions: Interference by cellular phone ringing occurred only with unprotected pacemaker models. Standard programming of these unprotected models was associated with a low incidence of interference.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2005.03.011DOI Listing

Publication Analysis

Top Keywords

cardiac pacemakers
12
cellular phone
12
electromagnetic filters
8
filters cardiac
8
cellular phones
8
output operating
8
mhz carrier
8
unprotected pacemaker
8
pacemaker models
8
incidence interference
8

Similar Publications

Implantable systems with chronic stability, high sensing performance, and extensive spatial-temporal resolution are a growing focus for monitoring and treating several diseases such as epilepsy, Parkinson's disease, chronic pain, and cardiac arrhythmias. These systems demand exceptional bendability, scalable size, durable electrode materials, and well-encapsulated metal interconnects. However, existing chronic implantable bioelectronic systems largely rely on materials prone to corrosion in biofluids, such as silicon nanomembranes or metals.

View Article and Find Full Text PDF

Introduction: The advancement of medical technology has introduced leadless pacemakers (LPMs) as a significant innovation in cardiac pacing, offering potential advantages over traditional ventricular transvenous pacemakers. This report explores the application of LPMs in two patients with complex valvular histories, particularly those with mechanical tricuspid valves.

Case Reports: The first case involves a 60-year-old male with a history of rheumatic heart disease and triple valve replacement who developed a high-grade AV block.

View Article and Find Full Text PDF

Prognostic role of aetiological agent vs. clinical pattern in candidates to lead extraction for cardiac implantable electronic device infections.

Sci Rep

December 2024

Department of Medical and Surgical Sciences, Institute of Cardiology, University of Bologna, Policlinico S.Orsola-Malpighi, via Massarenti 9, Bologna, 40138, Italy.

Cardiac implantable electronic devices infections (CIEDI) are associated with poor survival despite the improvement in transvenous lead extraction (TLE). Aetiology and systemic involvement are driving factors of clinical outcomes. The aim of this study was to explore their contribute on overall mortality.

View Article and Find Full Text PDF

A rare cardiac presentation of a lymphoma: case report.

Eur Heart J Case Rep

January 2025

Department of Cardiology, Klinik Floridsdorf, Brünnerstraße 68, Vienna 1210, Austria.

Background: Cardiac lymphoma is a rare disease that can present in various ways. Additionally, atypical clinical presentation makes the diagnosis even more challenging. The most common type of cardiac lymphoma is diffuse large B-cell lymphoma.

View Article and Find Full Text PDF

Objective: Phase angle (PhA) is a prognostic factor for predicting and monitoring geriatric syndromes. However, multiple factors associated with increased PhA values as an outcome remain unclear in the older population. This study aimed to examine the association of socio-demographic, anthropometric and behavioural factors with PhA among older Taiwanese adults.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!