Objective: To review the performance of polymethyl pentene versus silicone oxygenators in terms of efficiency in priming and oxygenation, oxygenator resistance, requirements for coagulation proteins and consumption of blood products, for neonatal extracorporeal membrane oxygenation (ECMO) patients.

Study Design: Forty consecutive neonates were selected retrospectively pre- and post-introduction of the new polymethyl pentene (PMP) oxygenators. They formed two equal groups. After calculation of the sample size, data were collected from ELSO registry forms and patient records. Results were analysed using parametric and non-parametric tests.

Results: Neonatal PMP (N-PMP) oxygenators were smaller, faster and easier to prime. They were less efficient than silicone oxygenators, especially in carbon dioxide elimination, and, therefore, required higher sweeps. The preservation of coagulation proteins was significantly better, but there was no reduction in the consumption of blood products, despite having less than half the surface area and significantly lower blood path resistance.

Conclusion: Small PMP oxygenators (Medos Hilite 800 LT) provide adequate gas exchange and offer technical advantages in terms of more efficient priming, reduced haemodynamic resistance and better control and preservation of coagulation proteins than silicone oxygenators.

Download full-text PDF

Source
http://dx.doi.org/10.1191/0267659105pf797oaDOI Listing

Publication Analysis

Top Keywords

polymethyl pentene
12
silicone oxygenators
12
coagulation proteins
12
performance polymethyl
8
oxygenators
8
neonatal extracorporeal
8
extracorporeal membrane
8
membrane oxygenation
8
consumption blood
8
blood products
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!