A method based on the presence of a dynamic pH junction within the capillary to induce band narrowing for enhanced detection sensitivity for some peptides is presented. This technique is predicated on a sharp reduction in an analyte's migration velocity following a reversal of its electrophoretic direction from the acidic sample zone to the basic BGS zone. Larger-than-usual injection volumes of samples in relatively high-conductivity matrices were enabled, without degrading peak shape, resolution and efficiency. The size of the original sample plug was reduced by as much as 38-fold, and improvement in detector response in terms of peak height by as much as 124-fold was obtained. The effects of pH and concentration of the sample matrix, and the length of sample injection on the efficiency of the technique are discussed.