A method based on the presence of a dynamic pH junction within the capillary to induce band narrowing for enhanced detection sensitivity for some peptides is presented. This technique is predicated on a sharp reduction in an analyte's migration velocity following a reversal of its electrophoretic direction from the acidic sample zone to the basic BGS zone. Larger-than-usual injection volumes of samples in relatively high-conductivity matrices were enabled, without degrading peak shape, resolution and efficiency. The size of the original sample plug was reduced by as much as 38-fold, and improvement in detector response in terms of peak height by as much as 124-fold was obtained. The effects of pH and concentration of the sample matrix, and the length of sample injection on the efficiency of the technique are discussed.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2005.03.069DOI Listing

Publication Analysis

Top Keywords

dynamic junction
8
junction technique
4
technique on-line
4
on-line preconcentration
4
preconcentration peptides
4
peptides capillary
4
capillary electrophoresis
4
electrophoresis method
4
method based
4
based presence
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!