Purpose: To compare the accuracy of biometry using conventional A-scan ultrasonography and partial coherence interferometry, and to improve the accuracy of biometry by sequential audit of postoperative refractive error.

Methods: The study was performed in three phases. In phase 1, 20 consecutive patients undergoing routine phacoemulsification underwent biometry using both A-scan ultrasonography and the Zeiss IOLMaster (ZIOLM). A single experienced optometrist refracted all patients 2 weeks after surgery. The errors between expected and achieved refraction were calculated and compared between the two methods. In phases 2 and 3, a further 22 and 20 patients, respectively, were recruited and only the ZIOLM was used for biometry. The manufacturer's suggested A-constant was refined and the error between expected and achieved refraction was calculated.

Results: In phase 1, the median unexpected error for the ZIOLM was+0.63 (interquartile range+0.368 to+1.015) and for A-scan biometry was --0.24 (interquartile range--1.335 to+0.802). In phase 1 65% of patients' postoperative refractions were found to be within 1.0 D of emmetropia using the ZIOLM (55% using A-scan biometry). Refinements to the A-constant improved this to 95% by phase 3.

Conclusion: An error was identified in IOL power estimation with the ZIOLM, when using the manufacturer's recommended A-constant (recommended and previously optimized ultrasound A-constant 118.0; ZIOLM optimized A-constant 118.6). Serial modifications to the A-constant were successful in reducing the unexpected error to well within the tolerance limits of published international standards.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1395-3907.2005.00486.xDOI Listing

Publication Analysis

Top Keywords

zeiss iolmaster
8
accuracy biometry
8
a-scan ultrasonography
8
expected achieved
8
achieved refraction
8
unexpected error
8
a-scan biometry
8
biometry
6
ziolm
6
a-constant
6

Similar Publications

Changes in Axial Length in Patients of Acute Retinal Necrosis Undergoing Vitrectomy.

Ocul Immunol Inflamm

January 2025

Eye Institute and Department of Ophthalmology, Eye and ENT Hospital, Fudan University, Shanghai, China.

Purpose: To observe the changes in the axial length (AL) in patients of acute retinal necrosis (ARN) undergoing vitrectomy and investigate the correlated factors.

Methods: Retrospective case series. Patients diagnosed as ARN undergoing vitrectomy with silicone oil (SO) tamponade, and with attached retina more than one year after silicone oil removal (SOR) were included.

View Article and Find Full Text PDF

Purpose: To apply a new method, the Taylor Diagram, and a new concept, the centered root mean square error (cRMSE), in evaluating the performance of IOL formulas.

Methods: The preoperative biometrics were measured with the IOLMaster 700 (Carl Zeiss Meditec) and the postoperative spherical equivalent refraction (SER) was calculated in 888 anonymous patients. The Taylor Diagram was applied to visualize the centered root mean square error (cRMSE) and the correlation coefficient between the predictions and the observations (R).

View Article and Find Full Text PDF

Purpose: To evaluate the accuracy of toric intraocular lens (IOL) axis prediction between two preoperative measurement devices: the optical biometry (IOLMaster 500 or 700) and the dual Scheimpflug topography (Galilei G4).

Methods: Medical records of 64 eyes from 44 patients who underwent phacoemulsification and posterior chamber toric IOL (Zeiss AT TORBI 709M) implantation between July 2017 and January 2022 were reviewed. All patients underwent preoperative evaluation by optical biometry (IOLMaster 500 or IOLMaster 700) and Galilei G4.

View Article and Find Full Text PDF

Purpose: To investigate different measures for corneal astigmatism in the context of reconstructed corneal astigmatism (recCP) as required to correct the pseudophakic eye, and to derive prediction models to map measured corneal astigmatism to recCP.

Methods: Retrospective single centre study of 509 eyes of 509 cataract patients with monofocal (MX60P) IOL. Corneal power measured with the IOLMaster 700 keratometry (IOLMK), and Galilei G4 keratometry (GK), total corneal power (TCP2), and Alpin's integrated front (CorT) and total corneal power (CorTTP).

View Article and Find Full Text PDF

Purpose: This study evaluates the discrepancies between ACV measurements obtained from the Heidelberg Anterion and Zeiss IOLMaster 700 and investigates the significance of ACV and other ocular biometry parameters.

Patients And Methods: To investigate intraocular fluid circulation, a robust formula was developed for ACV measurement using the Zeiss IOLMaster 700. A pilot study was conducted to validate this formula, which relied on WTW, CCT, and ACD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!