Background And Purpose: For some treatment sites, there is evidence in the literature that five to nine equi-angular input beam directions are enough for generating IMRT plans. For oesophagus cancer, there is a report showing that going from four to nine beams may even result in lower quality plans. In this paper, our previously published algorithm for automated beam angle selection (Cycle) has been extended to include segmented IMRT. For oesophagus cancer patients, we have investigated whether automated orientation selection from a large number of equi-angular input beam directions (up to thirty-six) for IMRT optimisation can result in improved lung sparing.

Materials And Methods: CT-data from five oesophagus patients treated recently in our institute were used for this study. For a prescribed mean PTV dose of 55 Gy, Cycle was used in an iterative procedure to minimise the mean lung dose under the following hard constraints: standard deviation for PTV dose inhomogeneity 2% (1,1 Gy), maximum spinal cord dose 45 Gy. Conformal radiotherapy (CFRT) and IMRT plans for a standard four field oesophagus beam configuration were compared with IMRT plans generated by automated selection from nine or thirty-six equi-angular input beam orientations. Comparisons were also made with dose distributions generated with our commercial treatment planning system (TPS), and with observations in the literature.

Results: Using Cycle, automated orientation selection from nine or thirty-six input beam directions resulted in improved lung sparing compared to the four field set-ups. Compared to selection from nine input orientations, selection from thirty-six directions did always result in lower mean lung doses, sometimes with even fewer non-zero weight beams. On average only seven beams with a non-zero weight were enough for obtaining the lowest mean lung dose, yielding clinically feasible plans even in case of thirty-six input directions for the optimisation process. With our commercial TPS we observed the same contra-intuitive, unfavourable results as reported in the literature; nine field equi-angular IMRT plans had substantially higher mean lung doses than plans for the conventional four field set-ups. For all cases, the Cycle plans generated from nine equi-angular input directions were superior compared to similar plans generated with our commercial TPS.

Conclusions: For the studied oesophagus cancer patients the best plans for IMRT were obtained with Cycle, using automated beam orientation selection from thirty-six input beam directions. The lowest mean lung doses could be obtained with, on average, a selection of only seven beams with non-zero weight.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2005.06.028DOI Listing

Publication Analysis

Top Keywords

input beam
20
equi-angular input
16
beam directions
16
imrt plans
16
selection thirty-six
16
oesophagus cancer
12
orientation selection
12
plans generated
12
thirty-six input
12
lung doses
12

Similar Publications

Study on a Strategy to Improve the Image Quality and Imaging Depth for Novel Synthetic Aperture Schemes: An Experimental Investigation.

Ultrason Imaging

January 2025

Biomedical Ultrasound Imaging Laboratory, Department of Applied Mechanics and Biomedical Engineering, Indian Institute of Technology, Madras, Chennai, India.

Imaging depth remains a restriction for Synthetic Aperture (SA) approaches, even though SA techniques have been shown to overcome some of the drawbacks of Conventional Focused Beamforming (CFB) technique. This limitation is attributed to lesser energy per transmit in SA techniques compared to that of CFB technique. Therefore, in this paper, a systematic investigation is done to evaluate the improvement in imaging depth and image quality of B-mode ultrasound images in the case of SA technique using PZT transducer by boosting the input voltage to the transducer, while measuring the acoustic exposure parameters recommended in international standards.

View Article and Find Full Text PDF

Particularly when they occur at high speeds, vehicle accidents represent a serious threat to human beings and due to this fact vehicle accident is considered as worlds high priority risk. Several research have been done to enhance the crashworthiness of bumper subsystems. With an emphasis on the major crash management system components which are also known as crash box and bumper beam, this study explores ways to improve the crashworthiness of vehicles.

View Article and Find Full Text PDF

Purpose: During endovascular revascularization interventions for peripheral arterial disease, the standard modality of X-ray fluoroscopy (XRF) used for image guidance is limited in visualizing distal segments of infrapopliteal vessels. To enhance visualization of arteries, an image registration technique was developed to align pre-acquired computed tomography (CT) angiography images and to create fusion images highlighting arteries of interest.

Methods: X-ray image metadata capturing the position of the X-ray gantry initializes a multiscale iterative optimization process, which uses a local-variance masked normalized cross-correlation loss to rigidly align a digitally reconstructed radiograph (DRR) of the CT dataset with the target X-ray, using the edges of the fibula and tibia as the basis for alignment.

View Article and Find Full Text PDF

Silica Waveguide Thermo-Optic Mode Switch with Bimodal S-Bend.

Nanomaterials (Basel)

December 2024

State Key Laboratory of Integrated Optoelectronics, College of Electronic Science & Engineering, Jilin University, No. 2699 Qianjin Street, Changchun 130012, China.

A silica waveguide thermo-optic mode switch with small radius bimodal S-bends is demonstrated in this study. The cascaded multimode interference coupler is adopted to implement the E and E mode selective output. The beam propagation method is used in design optimization.

View Article and Find Full Text PDF

Transmission matrix measurements of multimode fibers are now routinely performed in numerous laboratories, enabling control of the electric field at the distal end of the fiber and paving the way for the potential application to ultrathin medical endoscopes with high resolution. The same concepts are applicable to other areas, such as space division multiplexing, targeted power delivery, fiber laser performance, and the general study of the mode coupling properties of the fiber. However, the process of building an experimental setup and developing the supporting code to measure the fiber's transmission matrix remains challenging and time consuming, with full details on experimental design, data collection, and supporting algorithms spread over multiple papers or lacking in detail.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!