A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Evaluation of the MagNA Pure LC used with the TRUGENE HBV Genotyping Kit. | LitMetric

Evaluation of the MagNA Pure LC used with the TRUGENE HBV Genotyping Kit.

J Clin Virol

Division of Clinical Microbiology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN 55905, USA.

Published: October 2005

Background: The current manual sample processing method recommended for use with the TRUGENE HBV Genotyping Kit (TRUGENE HBV; Bayer HealthCare LLC, Tarrytown, NY) is labor-intensive and may be prone to specimen cross-contamination. Recent evaluations of the MagNA Pure LC (MP; Roche Applied Science, Indianapolis, IN) suggest that it is suitable for automated, contamination-free extraction and purification of viral nucleic acids from large-volume (1.0 mL) serum or plasma specimens.

Objectives: We evaluated the MP Total Nucleic Acid Isolation Kit--Large Volume (Roche Applied Science) in conjunction with TRUGENE HBV to establish the analytical sensitivity (threshold titer) of the assay, in HBV DNA International Units (IU)/mL, for obtaining consistent, interpretable sequence data from TRUGENE HBV.

Study Design: HBV analytical standards, prepared as 10 replicates (1.0 mL each) at each of the following concentrations: 200, 1000, 5000, and 10,000 IU/mL, were processed by MP and analyzed by TRUGENE HBV according to manufacturer's instructions. Performance of TRUGENE HBV used in conjunction with MP sample processing was evaluated further using 22 clinical serum specimens containing low titers of HBV DNA.

Results: All replicates of HBV analytical standards at 1000, 5000, and 10,000 IU/mL yielded interpretable TRUGENE HBV sequences, whereas interpretable sequences were obtained in 90% (9 of 10) of the replicates at 200 IU/mL. TRUGENE HBV sequences were interpretable in 86% (19 of 22) of the clinical specimens studied.

Conclusions: MP sample processing is efficient and suitable for use with TRUGENE HBV. When combined with MP sample processing, TRUGENE HBV yielded interpretable sequences from HBV analytical standards and clinical serum specimens with HBV DNA titers of > or =200 IU/mL.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcv.2005.05.008DOI Listing

Publication Analysis

Top Keywords

trugene hbv
40
hbv
16
sample processing
16
hbv analytical
12
analytical standards
12
trugene
11
magna pure
8
hbv genotyping
8
genotyping kit
8
roche applied
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!