Background And Purpose: To investigate the effect of radiotherapy planning with a dedicated combined PET-CT simulator of patients with locally advanced non-small cell lung cancer.

Patients And Methods: Twenty-one patients underwent a pre-treatment simulation on a dedicated hybrid PET-CT-simulator. For each patient, two 3D conformal treatment plans were made: one with a CT based PTV and one with a PET-CT based PTV, both to deliver 60Gy in 30 fractions. The maximum tolerable prescribed radiation dose for CT versus PET-CT PTV was calculated based on constraints for the lung, the oesophagus, and the spinal cord, and the Tumour Control Probability (TCP) was estimated.

Results: For the same toxicity levels of the lung, oesophagus and spinal cord, the dose could be increased from 55.2+/-2.0Gy with CT planning to 68.9+/-3.3Gy with the use of PET-CT (P=0.002), with corresponding TCP's of 6.3+/-1.5% for CT and 24.0+/-5.6% for PET-CT planning (P=0.01).

Conclusions: The use of a combined dedicated PET-CT-simulator reduced radiation exposure of the oesophagus and the lung, and thus allowed significant radiation dose escalation whilst respecting all relevant normal tissue constraints.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2005.06.014DOI Listing

Publication Analysis

Top Keywords

radiotherapy planning
8
planning dedicated
8
dedicated combined
8
non-small cell
8
cell lung
8
based ptv
8
radiation dose
8
lung oesophagus
8
oesophagus spinal
8
spinal cord
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!